4.3 Demo Mode
The purpose of Demo mode is to show a visual presentation demonstrating how the
N5500 monitor works. The following
pr
ocedure is set to Demo mode.
1.Turn the N5500 monitor on.
2.Rotate the knob to highlight the Set-up icon, then press the knob. Set-up menu
displays.
3.Rotate the knob to highlight Service Menu in the Set-up menu (see Figure 5), and
then press the knob to enter the Demo access code for Demo mode.
4.Three digits displays in the Level 2 Menu as shown in Figure 5.
5.Rotate the knob to highlight the top of the digits. Press the knob to enter the Demo
access code, “
3
” “
1
” “
4
” in this order.
6.The monitor is now set to Demo mode, and demonstrates a typical monitoring screen.
4.4 Factory Default Settings
Factory default settings are divided into three groups, adult, pediatric and neonatal as
described in Table 12.
The parameters of the N5500 monitor are preset to “Adult” monitoring mode. Alarm
limits settings will be automatically changed to the default settings for each patient
mode as the mode is changed to Adult, Pediatric or Neonatal mode.
Table 12. Factory Default Settings for the N5500
Factory Defaults
Parameter
Adult
Pediatric
Neonatal
Monitoring Mode
Adult
Pediatric
Neonatal
NIBP Initial Cuff Inflation
180 mmHg
180 mmHg
120 mmHg
NIBP Auto Mode Interval
OFF
OFF
OFF
NIBP SYS Alarm Upper Limits
160 mmHg
120 mmHg
90 mmHg
NIBP SYS Alarm Lower Limits
90 mmHg
70 mmHg
40 mmHg
NIBP DIA Alarm Upper Limits
90 mmHg
70 mmHg
60 mmHg
NIBP DIA Alarm Lower Limits
50 mmHg
40 mmHg
20 mmHg
NIBP MAP Alarm Upper Limits
110 mmHg
90 mmHg
70 mmHg
NIBP MAP Alarm Lower Limits
60 mmHg
50 mmHg
30 mmHg
ECG Size
10.0 mm/mV
10.0 mm/mV
10.0 mm/mV
ECG Filter Mode
Monitor
Monitor
Monitor
ECG Pacer Detect
OFF
OFF
OFF
ECG Sweep Speed
25.0 mm/s
25.0 mm/s
25.0 mm/s