Section 3: Performance Verification
3.1 Introduction
3.2 Equipment Needed
3.3 Performance Tests
3.4 Safety Tests
N5500 Service Manual 9
Performance
Verification, N5500
3.1 Introduction
This section discusses the tests used to verify performance following repairs or during
routine maintenance. All tests can be performed without removing the N5500 covers.
All tests except the battery charge and battery performance tests must be performed as
the last operation before the monitor is returned to the user.
If the N5500 fails to perform as specified in any test, repairs must be made to correct
the problem before the monitor is returned to the user.
3.2 Equipment Needed
Table 1 lists the equipment required for performance verification.
Table 1. Required Test Equipments
Equipment
Description
Digital multimeter (DMM)
Fluke Model 87 or equivalent
ECG cable for 3-leadwires
MDEC No.2
ECG cable for 5-leadwires (optional)
MDEC No.4
ECG leads
MDEL series (MDEL-3S, MDEL-3G)
ECG leads (optional)
MDEL series (MDEL-5S, MDEL-5G)
NIBP hose
Colin Hose No.1EM/No.3EM
NIBP cuff
Colin Cuff series
Pulse oximetry cable
DOC-10
O
XI
M
AX
Durasensor
® oxygen transducer DS-100A
O
XI
M
AX
oxygen transducer
MAX-A
Temperature probes
Monotherm™
SpO
2
simulator
Nellcor SRC-MAX simulator
ECG simulator
METRON PS-420 or equivalent
NIBP simulator
Bio-Tek “BP Pump 2” or equivalent
Respiration simulator
METRON PS-420 or equivalent
Temperature simulator
medSim 300 or equivalent
Safety analyzer
METRON QA-90 or equivalent
Data interface cable
RS-232 cable
Stopwatch
Manual or electronic