Section 7: Disassembly Guide
7-11
7.7
REPLACING THE LCD AND LED DISPLAY
(Refer to Figure 7-12: Replacing LCD and LED display)
7.7.1
Replacing the LCD
1.
Remove the CO
2
board as described in Figure 7-4: Replacing CO
2.
Unscrew and remove the four LCD mounting screws.
3.
Carefully remove the LCD and replace with a new LCD.
7.7.2
Replacing the LED Display
1.
Remove the CO
2
board as described in section 7.3 "Replacing the
2.
Carefully remove the faulty LED display from its socket and
replace with a new LED display. Note the correct orientation of
the LED display when replacing it.
Figure 7-12: Replacing LCD and LED display