SECTION 3: CLEANING
3.1
Cleaning
3.1
CLEANING
Always clean the outer surface of the monitor before servicing.
Warning: Always wear latex or surgical gloves when cleaning
or servicing the monitor after hospital use.
Warning: Do not allow liquids to drip inside the housing.
Warning: Do not clean internal components.
Warning: Do not use preparations based on: phenols, halogen
producing compounds, strong organic acids, or oxygen
producing compounds.
3.1.1
Procedure
1.
Unplug the unit from AC power.
2.
With a damp cloth, gently wipe the outer case of the monitor with a
disinfectant based on aldehyde, alcohol or quaternary ammonium
producing compounds.