6-3
Section 6: Flow Calibration Check
Objective
Action
Response
If Fail is displayed
Replace Flow
System and/or
Pump
(refer to Section 5:
Troubleshooting)
and repeat above
procedure.
If Pass is displayed
Exit Service Mode.
To reset and return to
measuring mode
long press
To exit service mode
and save event
messages
Turn off the
monitor.
Table 6-1: Flow Calibration (Continued)