DC-3000C SERIES ULTRASONIC THICKNESS GAUGE
- 11 -
Press
to confirm.
8.5.1.4 Velocity storage
This function allows user to store 4 new velocities Rate.
8.5.2
Resolution
Users can select the displayed resolution. When selecting the high
accuracy, the workpiece surface to be measured should be smooth, for
the purpose of getting a accurate value.
Press
key into
“Resolution”,
Press
or
to select resolution and unit.
1. 0.1 mm
2. 0.01 mm
3. 0.01 in
4. 0.001 in
Press
key to enter/confirm
Notice: When probe D5113 and D2012 is selected, it is recommended to
use 0.1 mm and 0.01in.
8.5.3
Probe Calibration
It will cause error during the primary stage of usage and operating, this
caused by the following three aspects:
1. The probe itself or the temperature variation
2. System error caused by the match of the instrument and the probe.