DC-3000C SERIES ULTRASONIC THICKNESS GAUGE
- 15 -
8.7.2
Gain setting
Press
into the
“Gain adjustment”
, the screen will display:
1. High
2. Medium
3. Low
4. Automatic
Press
or
to select desired item
Press
to confirm.
In the user's measuring environment, both different materials and the
same material with different status will have different effects on the
accurate and stable measuring. So for different measured objects and
different measuring environment, users should adjust the work status of
the instrument to meet more measurements.
For many materials and measuring conditions, auto gain adjustment can
be used, but for some special measurement, adjusting the instrument's
working status is necessary. There are four different working modes: Auto,
Low, medium and high.
Auto: match different probe and meets almost all the measuring
requirements.
Low:
Suitable for high scattering and small attenuation materials,
Medium:
Suitable for many measurements,
High:
Suitable for high attenuation material