DC-3000C SERIES ULTRASONIC THICKNESS GAUGE
- 8 -
Display the current value, satisfied with the normal measuring needs.
Minimum Measurement:
Among one measurement, display the point. It is suitable for testing the
curvature surface or needs to get the minimum value which is widely used
in the thickness measurement of pipeline.
Note: it is not recommended to use this function when measuring cast
iron or alloy materials.
Difference mode: (DC-3020C)
Display the accurate differential value between the measured value and
reference value set by the users, suitable for quality check to identifying
the qualified products, whose thickness is in the admissive error.
Average mode: (DC-3020C)
Provides the average value of 2 to 9 measured points and display it
suitable for testing the flat surface.
Limitation setting: (DC-3020C)
Set the upper and lower limit, when the measured thickness exceeds the
preset limit, it will display and give alarm, this measurement mode is more
widely used than differential mode.
Scan:
It is available for measuring the thickness of test piece with high
temperature surface. The gauge will alarm for each fast measurement.
And will display the all measured thickness upon the complete
measurement finished.
8.4.1.2 E-E Mode:
When the E-E mode is selected, only probe D5301 is available. This
function allows you to make measurement between two consecutive back