49
Isolations Transformer
The Isolation Transformer is intended for applications where medical devices require improved electrical isolation
and/or reduced leakage current to comply with existing safety standards.
This device has been designed and tested in accordance with IEC 60601-1 Medical Electrical Equipment standard.
To ensure a safe operation and to maintain the device in a safe condition, follow the instructions in this manual and pay
attention to all warnings and cautions.
Models: 9031D040x / 9031D041x
Function
Description
1
Mains Input
Plug the shielded mains power cord supplied with the system into this socket.
NOTE
Only use a shielded power cord from Natus.
Rating Frequency: 50/60Hz.
- Continuous Operation: max. 1000VA
- Momentary Operation: max 1500VA
9031D040x: 100/120V~
9031D041x: 200/240V~
2
Fuse
F1 and F2 Mains
9031D040x: 110V/120V: T10AH / 250V~
9031D041x: 200/240V: T6.3AH / 250V~
3
M4 Thread for Electrical Enclosure Contact and Additional Protective Earth.
4
Potential equalization
Where potential equalization is necessary or desirable, a potential equalization conductor may be
connected to this connector as per DIN 42 801.