IEC 60601-1-2:2007 and EN 60601-1-2:2007
Guidance and manufacturer's declaration - electromagnetic emissions for all equipment and systems
HI-PRO 2 is intended for use in the electromagnetic environment specified below. The user of HI-PRO 2 should
ensure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
HI-PRO 2 uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
HI-PRO 2 is suitable for use in all environments, including domestic
environments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic pur-
poses.
Guidance and manufacturer's declaration - electromagnetic immunity for all equipment and systems
HI-PRO 2 is intended for use in the electromagnetic environment specified below. The user of HI-PRO 2 should
ensure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment - guid-
ance
Electrostatic discharge
(ESD)
IEC 61000-4-2
+/- 6 kV contact
+/- 8 kV air
+/- 6 kV contact
+/- 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humid-
ity should be at least 30 %.
Electrical fast tran-
sient/burst IEC 61000-
4-4
+/- 1 kV for input/out-
put lines
No relevant ports
that could be
affected
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical com-
mercial or hospital environment.
Otometrics - HI-PRO 2
15
Содержание otometrics HI-PRO 2
Страница 1: ...HI PRO 2 Installation Manual Doc No 7 50 0980 EN 03 Part No 7 50 09800 EN...