Essential performance
HI-PRO 2 has no essential performance.
Standards
Safety
IEC 60601-1:2005+AMD1:2012
EN 60601-1:2006+A1:2013
ANSI/AAMI ES60601-1:2005 + A1:2012
CAN/CSA-C22.2 NO. 60601-1:14
EMC
IEC 60601-1-2:2007
EN 60601-1-2:2007
IEC 60601-1-2:2014
EN 60601-1-2:2015
Systems
IEC 60601-1:2005+AMD1:2012
EN 60601-1:2006+A1:2013
Hearing Instrument
Interface
IEC 60118-14:1998
EN 60118-14:1998
To comply with the above standards, the programming cable and the connector to
the hearing instrument must meet the following requirements:
•
No conductive parts may be accessible when the programming cable is con-
nected to HI-PRO 2 unit.
•
The programming cable and the connector must provide double isolation and be
able to withstand a dielectric strength test potential of 500 V.
Output ratings
The following output ratings are valid for Left and Right hearing instrument con-
nectors.
Fixed Battery supply (pin1),
common for Left and Right side
1.35V, 10/50 mA (current rating is controlled by fit-
ting software)
Programmable battery supply
(pin5), common for Left and
Right side
-3.50 V to +3.50 V, 30 mA (voltage is controlled by
fitting software)