background image

Portable and mobile high-frequency communication devices, such as mobile
phones, may interfere with the functioning of HI-PRO 2.

IEC 60601-1-2:2014 and EN 60601-1-2:2015

Guidance and manufacturer's declaration - electromagnetic emissions for all equipment and systems

HI-PRO 2 is intended for use in the electromagnetic environment specified below. The user of HI-PRO 2 should

ensure that it is used in such an environment.

Emissions test

Compliance

Electromagnetic environment - guidance

RF emissions

CISPR 11

Group 1

HI-PRO 2 uses RF energy only for its internal function. Therefore, its RF

emissions are very low and are not likely to cause any interference in

nearby electronic equipment.

RF emissions

CISPR 11

Class B

HI-PRO 2 is suitable for use in all environments, including domestic

environments and those directly connected to the public low-voltage

power supply network that supplies buildings used for domestic pur-

poses.

Guidance and manufacturer's declaration - electromagnetic immunity for all equipment and systems

HI-PRO 2 is intended for use in the electromagnetic environment specified below. The user of HI-PRO 2 should

ensure that it is used in such an environment.

Immunity test

IEC 60601

test level

Compliance level

Electromagnetic environment - guid-

ance

Electrostatic discharge

(ESD)

IEC 61000-4-2

+/- 8 kV contact

+/- 2 kV, +/- 4 kV,

+/- 8 kV, +/- 15 kV air

+/- 8 kV contact

+/- 2 kV, +/- 4 kV,

+/- 8 kV, +/- 15 kV air

Floors should be wood, concrete or

ceramic tile. If floors are covered with

synthetic material, the relative humid-

ity should be at least 30 %.

Electrical fast tran-

sient/burst IEC 61000-

4-4

+/- 1 kV for input/out-

put lines

No relevant ports

that could be

affected

Power frequency

(50/60 Hz) magnetic

field

IEC 61000-4-8

30 A/m

No relevant ports

that could be

affected

Power frequency magnetic fields

should be at levels characteristic of a

typical location in a typical com-

mercial or hospital environment.

Otometrics - HI-PRO 2

13

Содержание otometrics HI-PRO 2

Страница 1: ...HI PRO 2 Installation Manual Doc No 7 50 0980 EN 03 Part No 7 50 09800 EN...

Страница 2: ...d Otometrics the Otometrics Icon AURICAL MADSEN Otoscan ICS and HORTMANN are registered trademarks of Natus Medical Denmark ApS in the U S A and or other countries Version release date 2019 02 05 2068...

Страница 3: ...ble of Contents 1 Introduction to HI PRO 2 4 2 When you receive HI PRO 2 5 3 Installation 6 4 Service and Maintenance 9 5 Technical Specifications 10 6 Safety 18 7 Manufacturer 20 Otometrics HI PRO 2...

Страница 4: ...les for connecting hearing instruments to HI PRO 2 are supplied by the hearing instrument manufacturer 1 1 Intended use HI PRO 2 is intended for audiologists hearing instrument dispensers and other he...

Страница 5: ...ient or risk of damage to data or the device Note Indicates that you should take special notice 2 When you receive HI PRO 2 2 1 Unpacking and inspection 1 Unpack your HI PRO carefully When you unpack...

Страница 6: ...ollow the guidelines below Store HI PRO 2 and accessories in the box provided to protect the equipment from damage Store HI PRO 2 as stated in Technical Specifications 3 Installation Site the HI PRO 2...

Страница 7: ...y double clicking the icon on the desktop double click on the CD drive icon then double click on the Setup application icon to start the installation Follow the instructions on the screen 3 2 Connecti...

Страница 8: ...icate communication with the PC Caution Do not attempt to connect or disconnect a hearing instrument while the connector LED is active The LED indicates that the connector is act ive and this might da...

Страница 9: ...o circumstances disassemble HI PRO 2 Contact your supplier Parts inside HI PRO 2 must only be checked or serviced by authorized personnel Warning Do not disassemble the HI PRO 2 as there is a risk of...

Страница 10: ...port is used for communication between a PC and the HI PRO 2 unit Communication USB 2 0 full Speed USB 1 1 compatible USB Connector Type Type B connector on the HI PRO 2 unit Power Supply The HI PRO...

Страница 11: ...hearing instrument must meet the following requirements No conductive parts may be accessible when the programming cable is con nected to HI PRO 2 unit The programming cable and the connector must pr...

Страница 12: ...t Size L x W x H 137 mm x 114 mm x 37 mm 5 39 inch x 4 49 inch x 1 46 inch Net weight 230g 0 43 lb 5 1 Accessories Item Part number 1072 HI PRO 2 Installation CD 8 49 91200 USB Cable 3 meters 8 71 791...

Страница 13: ...ower supply network that supplies buildings used for domestic pur poses Guidance and manufacturer sdeclaration electromagnetic immunity for all equipment and systems HI PRO 2 is intended for use in th...

Страница 14: ...GHz 10 V m 80 MHz to 2 7 GHz Proximity fieldsfrom RF wirelesscom munications IEC 61000 4 3 27 V m 386 MHz 28 v m 450 MHz 9 v m 710 MHz 745 MHz 780 MHz 28 v m 810 MHz 870 MHz 930 MHz 28 v m 1720 MHz 18...

Страница 15: ...lies buildings used for domestic pur poses Guidance and manufacturer sdeclaration electromagnetic immunity for all equipment and systems HI PRO 2 is intended for use in the electromagnetic environment...

Страница 16: ...HI PRO 2 including cables than the recommended separation distance cal culated from the equation applicable to the frequency of the transmitter Recommended separation distance d 1 2 d 1 2 for 80 MHzt...

Страница 17: ...e is observed additional measures might be necessary such as reorienting or relocating HI PRO 2 b Over the frequency range 150 kHz to 80 MHz field strengths should be less than 3 V m Recommended separ...

Страница 18: ...1 2005 AMD1 2012 and EN 60601 1 2006 A1 2013 Class II equipment Follow instructions for use Consult user manual for warnings and cautions Complies with Medical Devices Directive 93 42 EEC and RoHS Di...

Страница 19: ...s When connecting equipment to the USB connector the following must be considered Equipment must be certified to relevant EN IEC safety stand ards e g EN IEC 60950 Use of connected equipment in a pati...

Страница 20: ...f HI PRO 2 6 No parts may be eaten burnt or in any way used for purposes other than the fitting of hearing aids or similar devices 7 For safety reasons and due to effects on EMC accessories connected...

Страница 21: ...arried out by the equipment manufacturer or personnel authorized by the man ufacturer The electrical installation to which the equipment is connected complies with EN IEC requirements The equipment is...

Страница 22: ......

Отзывы: