Performing an Exam
Example Motor Nerve Conduction Test
Position and secure the electrodes to your patient according to your conventions
for the type of test you are performing.
Note:
The recording electrode is placed on a muscle, even though the interest is
in the conduction along the nerve.
The nerve is stimulated using surface electrodes at two or more sites where the
nerve is located superficially. The stimulator is oriented so that the cathode of the
stimulator faces the active recording electrode.
Negative voltage applied to the black (-) input jack produces an upward
deflection on the screen display. Use the black input jack as the
active
electrode
and the red input jack as the
reference
or
inactive
electrode for a negative-up
convention.
Electrode
Amplifier Channel Connector
Ground
Green DIN Input
Black (-) - Active Recording
Black DIN Input
Red (+) - Reference Recording
Red DIN Input
DIN (Recording)
Large 5-pin Circular DIN Input
Содержание Nicolet Viking EDX
Страница 2: ......
Страница 10: ...Nicolet EDX h 30 January 2020 Blank page...
Страница 15: ...January 30 2020 1 1 1 Introduction...
Страница 16: ...Nicolet EDX 1 2 January 30 2020 Blank page...
Страница 22: ...Nicolet EDX 2 2 January 30 2020 Blank Page...
Страница 54: ...Nicolet EDX 3 2 January 30 2020 Blank page...
Страница 58: ...Nicolet EDX 3 6 January 30 2020 Blank page...
Страница 60: ...Nicolet EDX 4 2 January 30 2020 Blank page...
Страница 72: ...Nicolet EDX 4 14 January 30 2020 Blank page...
Страница 74: ...Nicolet EDX 5 2 January 30 2020 Blank page...
Страница 86: ...Nicolet EDX 5 14 January 30 2020 Blank page...
Страница 88: ...Nicolet EDX 6 2 January 30 2020 Blank page...
Страница 92: ...Nicolet EDX 6 6 January 30 2020 Blank page...