94
NATIONAL
INSTRUMENTS
*STB?
Status Byte query. Returns the contents of the Status Byte Register.
Meaning and Type
Status Byte Device Status
Query Syntax
*STB?
Returned Parameters
<NR1> (Register binary value)
Bit Position
7
6
5
4
3
2
1
0
Condition
OPER
MSS
ESB
MAV
QUES
0
0
0
(RQS)
Bit Weight
128
64
32
16
8
4
2
1
ESB = Event status byte summary; MAV = Message available
MSS = Master status summary; OPER = Operation status summary;
QUES = Questionable status summary; RQS = Request for service
Table 7-11: Bit Configuration of Status Byte Register
*TRG
The Trigger command starts the waveform when the trigger source is set to BUS.
Meaning and Type
Trigger
Device Trigger
Command Syntax
*TRG
Parameters
None
Query Syntax
None
ABORt
Resets the trigger system and places the power supply in an IDLE state without waiting for the
completion of the trigger cycle.
Command Syntax
ABORt
Parameters
None
Example
ABOR
Query Syntax
None