NOTE:
VOLT:MODE LIST and WAVE is an implied ABORT command. WAVE mode cannot be programmed
simultaneously for both Voltage and Current. Only the last sent command can be accepted as WAVE.
Previous mode reverts to NONE.
VOLTage:PROTection:LEVel
Sets the OVP level. The OVP setting range is given in Table 7-9. The number of characters after OVP
is up to 12. The minimum setting level is approx. 105% of the set output voltage, or the value in
Table 7-9, whichever is higher.
SCPI Command Syntax
[SOURce]:VOLTage:PROTection:LEVel <nn.nn|MAX>
GEN Command Syntax
OVP < NRf+>
Default Suffix
V
*RST Value
MAX
Examples
VOLT:PROT:LEV 2.5
Query Syntax
[SOURce]:VOLTage:PROT:LEV?
VOLT:PROT:LEVel? MIN VOLT:PROT:LEVel? MAX
Returned Parameters
<NR3> VOLT:PROT:LEV? returns presently programmed OVP level. VOLT:PROT:LEV?
MAX and VOLT:PROT:LEV? MIN return the maximum and minimum programmable
OVP levels.
NOTE:
VOLT:PROT:LEV MIN set OVP level 5% above voltage set value
VOLTage:PROTection:LOW:STATe
Sets the under voltage protection (UVP) status of the power supply. If the UVP status selected,
then the under voltage protection is enabled.
SCPI Command Syntax
[SOURce]:VOLTage:PROTection:LOW:STATe <CRD>
GEN Command Syntax
UVL < NRf+> , UVP < NRf+>
Parameters
UVL|UVP
*RST Value
UVL
Examples
VOLT:PROT:LOW:STAT UVP
Query Syntax
[SOURce]:VOLTage:PROTection:LOW :STATe?
Returned Parameters
UVP|UVL
Содержание RMX-410 Series
Страница 1: ...RMX 4101...
Страница 30: ...29 NATIONAL INSTRUMENTS 2 4 1 Rackmount Dimensions Fig 2 3...
Страница 31: ...30 NATIONAL INSTRUMENTS Fig 2 4 Rackmount Dimensions Continued Fig 2 5 Power Supply Rear Support Bracket...
Страница 128: ...127 NATIONAL INSTRUMENTS...
Страница 130: ...NOTES...
Страница 131: ...NOTES...
Страница 132: ...NOTES...