background image

Digital I/O

The NI PXIe-7822R provides connections for 128 digital input/output (DIO) channels.
Connectors 0 through 3 each contain 32 high-speed DIO channels that can run up to 80 MHz
signal frequencies. Each connector has selectable logic levels that you can configure as 1.2 V,
1.5 V, 1.8 V, 2.5 V, or 3.3 V. You can configure each channel as input or output.

Figure 4.

 Connecting to the DIO Channels

NI PXIe-7822R

Power

FPGA

Connection

Accessory

DIO0

DIO1

DIO30

DIO31

Connector X (DIO)

1

2

1. High-speed signal frequencies up to 80 MHz with logic levels configured as 1.2 V, 1.5 V, 1.8 V, 2.5 V, or

3.3 V

2. LED

The DIO channels connect to the FPGA through protection circuitry, which has overvoltage
and undervoltage protection as well as overcurrent protection. Refer to the 

NI PXIe-7822R

Specifications

 for more information about the maximum voltage and current.

When the system powers on, the DIO channels are set as input low with pull-down resistors.
To set another power-on state, you can configure the NI PXIe-7822R to load a VI when the
system powers on. The VI can then set the DIO lines to any power-on state.

All the DIO channels on Connectors 0 through 3 are routed with a 50 Ω characteristic trace
impedance. Route all external circuitry with a similar impedance to ensure best signal quality.
NI recommends performing signal integrity measurements to test the affect of signal routing
with the cable and connection accessory for your application.

Installing Noise Suppression Ferrites

Caution

 

To ensure the specified EMC performance, operate this product only with

shielded cables and accessories. Do not use unshielded cables or accessories unless
they are installed in a shielded enclosure with properly designed and shielded input/
output ports and connected to the product using a shielded cable. If unshielded

8

  | 

ni.com

  | 

NI PXIe-7822R Getting Started Guide

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Содержание PXIe-7822R

Страница 1: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...

Страница 2: ...oduct misuse can result in a hazard You can compromise the safety protection built into the product if the product is damaged in any way If the product is damaged return it to National Instruments for...

Страница 3: ...lded cable If unshielded cables or accessories are not properly installed and shielded the EMC specifications for the product are no longer guaranteed Caution To ensure the specified EMC performance i...

Страница 4: ...ni com manuals for complete specifications Installing Software on the Host Computer Before using the NI PXIe 7822R you must install the following application software and device drivers on the host co...

Страница 5: ...nt panel 5 Identify a supported slot in the chassis The following figure shows the symbols that indicate the slot types Figure 2 Chassis Compatibility Symbols NI PXIe 1062Q 1 2 3 4 5 1 PXI Express Sys...

Страница 6: ...imize cooling air flow 12 Power on the chassis Verifying Hardware Installation for Host Targets You can verify that the system recognizes the NI PXIe 7822R by using Measurement Automation Explorer MAX...

Страница 7: ...t appear press F5 to refresh the view in MAX If the device does not appear after refreshing the view visit ni com support for troubleshooting information Connecting the NI PXIe 7822R NI recommends usi...

Страница 8: ...24 59 25 60 26 61 27 62 28 63 29 64 30 65 31 66 32 67 33 68 34 x is the connector number External Clock x is an input only DIO 0 31 Digital I O data through channels 0 through 31 GND Ground reference...

Страница 9: ...ers on the DIO channels are set as input low with pull down resistors To set another power on state you can configure the NI PXIe 7822R to load a VI when the system powers on The VI can then set the D...

Страница 10: ...ble install two 2 snap on ferrite beads 777297 01 one on each end of the cable as close to the connector as practical Two 2 snap on ferrite beads 777297 01 are included in ach of the following recomme...

Страница 11: ...everything from troubleshooting and application development self help resources to email and phone assistance from NI Application Engineers National Instruments corporate headquarters is located at 1...

Страница 12: ...le for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the National Instruments global trade compliance policy and how to obtain relevant HTS codes ECC...

Страница 13: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...

Отзывы: