Installing the NI PXIe-7822R

Caution

 

To prevent damage to the device caused by ESD or contamination, handle

the device using the edges or the metal bracket.

1.

Ensure the AC power source is connected to the chassis before installing the modules.

The AC power cord grounds the chassis and protects it from electrical damage while you
install the modules.

2.

Power off the chassis.

3.

Inspect the slot pins on the chassis backplane for any bends or damage prior to
installation. Do not install a module if the backplane is damaged.

4.

Remove the black plastic connectors from all the captive screws on the module front
panel.

5.

Identify a supported slot in the chassis. The following figure shows the symbols that
indicate the slot types.

Figure 2.

 Chassis Compatibility Symbols

NI PXIe-1062Q

1

2

3

4

5

1. PXI Express System Controller Slot
2. PXI Peripheral Slot
3. PXI Express Hybrid Peripheral Slot

4. PXI Express System Timing Slot
5. PXI Express Peripheral Slot

NI PXIe-7822R modules can be placed in PXI Express peripheral slots, PXI Express
hybrid peripheral slots, or PXI Express system timing slots.

6.

Touch any metal part of the chassis to discharge static electricity.

7.

Ensure that the ejector handle is in the unlatched (downward) position.

8.

Place the module edges into the module guides at the top and bottom of the chassis. Slide
the device into the slot until it is fully inserted.

4

  | 

ni.com

  | 

NI PXIe-7822R Getting Started Guide

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Содержание PXIe-7822R

Страница 1: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...

Страница 2: ...oduct misuse can result in a hazard You can compromise the safety protection built into the product if the product is damaged in any way If the product is damaged return it to National Instruments for...

Страница 3: ...lded cable If unshielded cables or accessories are not properly installed and shielded the EMC specifications for the product are no longer guaranteed Caution To ensure the specified EMC performance i...

Страница 4: ...ni com manuals for complete specifications Installing Software on the Host Computer Before using the NI PXIe 7822R you must install the following application software and device drivers on the host co...

Страница 5: ...nt panel 5 Identify a supported slot in the chassis The following figure shows the symbols that indicate the slot types Figure 2 Chassis Compatibility Symbols NI PXIe 1062Q 1 2 3 4 5 1 PXI Express Sys...

Страница 6: ...imize cooling air flow 12 Power on the chassis Verifying Hardware Installation for Host Targets You can verify that the system recognizes the NI PXIe 7822R by using Measurement Automation Explorer MAX...

Страница 7: ...t appear press F5 to refresh the view in MAX If the device does not appear after refreshing the view visit ni com support for troubleshooting information Connecting the NI PXIe 7822R NI recommends usi...

Страница 8: ...24 59 25 60 26 61 27 62 28 63 29 64 30 65 31 66 32 67 33 68 34 x is the connector number External Clock x is an input only DIO 0 31 Digital I O data through channels 0 through 31 GND Ground reference...

Страница 9: ...ers on the DIO channels are set as input low with pull down resistors To set another power on state you can configure the NI PXIe 7822R to load a VI when the system powers on The VI can then set the D...

Страница 10: ...ble install two 2 snap on ferrite beads 777297 01 one on each end of the cable as close to the connector as practical Two 2 snap on ferrite beads 777297 01 are included in ach of the following recomme...

Страница 11: ...everything from troubleshooting and application development self help resources to email and phone assistance from NI Application Engineers National Instruments corporate headquarters is located at 1...

Страница 12: ...le for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the National Instruments global trade compliance policy and how to obtain relevant HTS codes ECC...

Страница 13: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...

Отзывы: