Connector
Description
Connector Type
Q-
Input connector for Q-
signals.
MMPX (f)
Table 9.
LED Indicators
LED
Indications
ACCESS
Indicates the basic hardware status of the
device.
Off—The device is not yet functional or has
detected a problem with a PXI Express power
rail.
Amber—The device is being accessed. Accessed
means that you are writing to the device setup
registers to control the device, reading from the
device to monitor the device status, or
transferring data to/from the device.
Green—The device is controllable through the
software.
ACTIVE
Off—The device is idle.
Solid green—The device is generating a
waveform.
Dim amber—The device is waiting for an
acquisition Reference Trigger.
Solid amber—The device is acquiring a
waveform.
Solid red—The device has detected an error. The
LED remains red until the error condition is
removed.
Note
The indicators are listed in
increasing order of priority. For
example, if you are generating a
waveform using NI-RFSG and waiting
on an acquisition Reference Trigger
in NI-RFSA, the LED is dim amber.
© National Instruments
31
PXIe-5832 Getting Started