1.

Use the 2.92 mm-to-2.92 mm cable to connect the RF OUT connector on the NI 5698
front panel to the RF IN connector on the NI 5606 front panel.

2.

Hand-tighten all 2.92 mm cable ends on the 2.92 mm cable connectors. The cable
connectors should tighten without much torque or effort.

3.

Carefully complete tightening all 2.92 mm connectors to 1 N · m using an appropriate
torque wrench or torque screwdriver and driver bit.

Caution

 

Incorrect torque at 2.92 mm connections can degrade signal fidelity

and phase-locked loop (PLL) performance and may cause insertion loss. Use an
appropriate torque wrench or torque screwdriver and 2.92 mm driver bit to
ensure all 2.92 mm connections are properly torqued. 2.92 mm connectors for
connections to external equipment may require torque different from 1 N · m,
depending on the connector type, material, and manufacturer.

4.

Power on your PXI Express chassis and controller system.

5.

Verify that the ACCESS LED is illuminated on all modules.

Related Information

Troubleshooting

 on page 10

If an issue persists after you complete a troubleshooting procedure, contact NI technical
support or visit 

ni.com/support

.

Configuring the NI 5698 in MAX

 on page 9

NI PXIe-5698 Getting Started Guide

  | 

© National Instruments

  | 

7

Содержание PXIe-5698

Страница 1: ...t may be impaired if it is used in a manner not described in this document Hot Surface If the NI 5698 has been in use it may exceed safe handling temperatures and cause burns Allow the NI 5698 to cool...

Страница 2: ...ry rules Caution To ensure the specified EMC performance operate this product only with shielded cables and accessories Caution To ensure the specified EMC performance the length of all I O cables mus...

Страница 3: ...river Software DVD 6 NI PXIe 5698 Getting Started Guide This Document 7 Read Me First Safety and Electromagnetic Compatibility 8 Maintain Forced Air Cooling Note to Users Preparing the Environment Ens...

Страница 4: ...ion Accept the prompts to complete the installation 4 When the installer completes select Restart in the dialog box that prompts you to restart shut down or restart later Installing the NI 5698 Cautio...

Страница 5: ...p and bottom of the chassis Slide the device into the slot until it is fully inserted Figure 2 Module Installation 2 3 NI PXIe 1075 1 1 Chassis 2 Hardware Module 3 Ejector Handle in Downward Unlatched...

Страница 6: ...devices are maintained in a static free environment If you are using an active device such as a preamplifier or switch routed to the NI 5698 RF IN connector ensure that the device cannot generate sign...

Страница 7: ...LL performance and may cause insertion loss Use an appropriate torque wrench or torque screwdriver and 2 92 mm driver bit to ensure all 2 92 mm connections are properly torqued 2 92 mm connectors for...

Страница 8: ...evice Front Panel Icon Definitions Refer to the user documentation for required maintenance measures to ensure user safety and or preserve the specified EMC performance The signal pins of this product...

Страница 9: ...written to in order to control the device GREEN The module is ready to be programmed by NI RFSA ACTIVE Indicates the module state OFF The module is in a quiescent state AMBER The module is waiting for...

Страница 10: ...one device a Select the system device with the appropriate downconverter in the configuration tree b In the Associated Devices section use the RF Conditioning drop down listbox to select the NI 5698 c...

Страница 11: ...In the MAX configuration tree expand Devices and Interfaces 2 Expand the Chassis tree to see the list of installed devices and press F5 to refresh the list 3 If the module is still not listed power o...

Страница 12: ...ignal Analyzers Help NI PXIe 5698 Specifications NI RFSA Soft Front Panel NI RFSA Instrument Driver NI RF Vector Signal Analyzers Help NI RFSA Examples Located using the NI Example Finder Located at n...

Страница 13: ...the DoC for your product by visiting ni com certification If your product supports calibration you can obtain the calibration certificate for your product at ni com calibration NI corporate headquart...

Страница 14: ...party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy and how to obtain rel...

Отзывы: