12. Power on the chassis.
Related Information
on page 4
Direct Connections to the NI 5698
The NI 5698 is a precision RF instrument that is sensitive to ESD and transients. Ensure you
take the following precautions when making direct connections to the NI 5698 to avoid
damaging the device.
Caution
Apply external signals only while the NI 5698 is powered on. Applying
external signals while the device is powered off may cause damage.
•
Ensure you are properly grounded when manipulating cables or antennas connected to the
NI 5698 RF IN connector.
•
If you are using noninsulated devices, such as a noninsulated RF antenna, ensure the
devices are maintained in a static-free environment.
•
If you are using an active device, such as a preamplifier or switch routed to the NI 5698
RF IN connector, ensure that the device cannot generate signal transients greater than the
RF and DC specifications of the NI 5698 RF IN connector.
Connecting to the NI 5668R
Figure 3.
NI 5668R with NI 5698 Front Panel Interconnection
0 VDC
+ 20 dBm MAX > 10 MHz
+10 dBM MAX < 10 MHz
NOISE
SOURCE
0/28 VDC
ACCESS
ACTIVE
NI PXIe-5698
RF Preamplifier
10 MHz – 26.5 GHz
RF IN
+17 dBm MAX
±10 VDC
RF OUT
+25 dBm MAX
0 VDC
ACCESS
ACTIVE
ESD
SENSITIVE
ACCESS ACTIVE
LO3
800 MHz
LO2
4 GHz
LO1
3.2 GHz - 8.3 GHz
OUT
ALL PORTS
50
Ω
NI PXIe-5653
Synthesizer
REF IN
10 MHz
5 V p-p MAX
REF OUT
10 MHz
1.5 V p-p MAX
REF OUT
100 MHz
1.5 V p-p MAX
1
1. Semi-rigid 2.92 mm-to-2.92 mm Coaxial Cable
6
|
ni.com
|
NI PXIe-5698 Getting Started Guide