background image

Table 2.

 NI 5698 RF Preamplifier Module Front Panel Connectors

Connector

Use

 

 

NOISE SOURCE HD BNC connector for enabling noise source on or off to facilitate noise

figure measurements.

RF IN

Connects the analog RF input signal to the NI 5698 RF preamplifier. The
maximum RF input is +17 dBm.

RF OUT

Outputs the bypassed or preamplified RF signal. The maximum RF
output is +25 dBm.

Table 3.

 NI 5698 RF Preamplifier Module Front Panel LEDs

LED

Indications

 

 

ACCESS Indicates the basic hardware status of the module.

OFF—The module is not yet functional, or the module has detected a problem
with a PXI Express power supply rail.

AMBER—The module is being accessed. 

Accessed

 means that the device setup

registers are being written to in order to control the device.

GREEN—The module is ready to be programmed by NI-RFSA.

ACTIVE Indicates the module state.

OFF— The module is in a quiescent state.

AMBER—The module is waiting for a Reconfiguration Trigger from the
configuration list.

GREEN—The module is triggered and is running a step from the configuration
list.

RED—The module has detected an error state. An error state may indicate the
module has exceeded approved operating temperature and thermal shutdown has
occurred or that the module has detected a power supply failure. If the power
supply fails, contact NI technical support.

Configuring the NI 5698 in MAX

Use Measurement & Automation Explorer (MAX) to configure your NI hardware. MAX
informs other programs about which devices reside in the system and how they are configured.
MAX is automatically installed with NI-RFSA.

NI PXIe-5698 Getting Started Guide

  | 

© National Instruments

  | 

9

Содержание PXIe-5698

Страница 1: ...t may be impaired if it is used in a manner not described in this document Hot Surface If the NI 5698 has been in use it may exceed safe handling temperatures and cause burns Allow the NI 5698 to cool...

Страница 2: ...ry rules Caution To ensure the specified EMC performance operate this product only with shielded cables and accessories Caution To ensure the specified EMC performance the length of all I O cables mus...

Страница 3: ...river Software DVD 6 NI PXIe 5698 Getting Started Guide This Document 7 Read Me First Safety and Electromagnetic Compatibility 8 Maintain Forced Air Cooling Note to Users Preparing the Environment Ens...

Страница 4: ...ion Accept the prompts to complete the installation 4 When the installer completes select Restart in the dialog box that prompts you to restart shut down or restart later Installing the NI 5698 Cautio...

Страница 5: ...p and bottom of the chassis Slide the device into the slot until it is fully inserted Figure 2 Module Installation 2 3 NI PXIe 1075 1 1 Chassis 2 Hardware Module 3 Ejector Handle in Downward Unlatched...

Страница 6: ...devices are maintained in a static free environment If you are using an active device such as a preamplifier or switch routed to the NI 5698 RF IN connector ensure that the device cannot generate sign...

Страница 7: ...LL performance and may cause insertion loss Use an appropriate torque wrench or torque screwdriver and 2 92 mm driver bit to ensure all 2 92 mm connections are properly torqued 2 92 mm connectors for...

Страница 8: ...evice Front Panel Icon Definitions Refer to the user documentation for required maintenance measures to ensure user safety and or preserve the specified EMC performance The signal pins of this product...

Страница 9: ...written to in order to control the device GREEN The module is ready to be programmed by NI RFSA ACTIVE Indicates the module state OFF The module is in a quiescent state AMBER The module is waiting for...

Страница 10: ...one device a Select the system device with the appropriate downconverter in the configuration tree b In the Associated Devices section use the RF Conditioning drop down listbox to select the NI 5698 c...

Страница 11: ...In the MAX configuration tree expand Devices and Interfaces 2 Expand the Chassis tree to see the list of installed devices and press F5 to refresh the list 3 If the module is still not listed power o...

Страница 12: ...ignal Analyzers Help NI PXIe 5698 Specifications NI RFSA Soft Front Panel NI RFSA Instrument Driver NI RF Vector Signal Analyzers Help NI RFSA Examples Located using the NI Example Finder Located at n...

Страница 13: ...the DoC for your product by visiting ni com certification If your product supports calibration you can obtain the calibration certificate for your product at ni com calibration NI corporate headquart...

Страница 14: ...party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy and how to obtain rel...

Отзывы: