![National Instruments PXIe-4468 Скачать руководство пользователя страница 37](http://html1.mh-extra.com/html/national-instruments/pxie-4468/pxie-4468_calibration-procedure_3668256037.webp)
PXIe-4468 Calibration Procedure
© National Instruments Corporation
35
Call DAQmxInitExtCal (DAQmx Initialize External Calibration VI) to
initialize the calibration using the following parameters:
o
deviceName
:
Dev_name
o
password
: NI
o
calHandle
: &myCalHandle
2.
Connect the calibrator to AI0 as shown in
or
3.
Configure the calibrator to generate an output with the amplitude of
6.3 Vrms and frequency of 1kHz.
4.
Call DAQmxAdjust4468Cal (DAQmx Adjust 4468 Calibration VI from the
DAQmx Adjust DSA AI Calibration Polymorphic AI), using the following
parameters:
calHandle
: myCalHandle
referenceVoltage
: 6.3
Timebase Adjustment
Perform an adjustment at least once within the calibration interval.
Adjustment automatically updates the calibration constants, the date, and
the temperature in the DUT EEPROM. If the DUT passes the verification
procedures within the As-Left test limits, an adjustment is not required.
Proceed to the
Update the Onboard Calibration Information
section.
Note
Initial test connection is same as Timebase Frequency Verification, as
shown in
or
Adjustment Procedure
1.
Connect channel AI0 to the function generator as shown in
Содержание PXIe-4468
Страница 1: ...PXIe 4468...
Страница 2: ...September 2022 PXIe 4468 Calibration procedure...
Страница 3: ...ni com services PXIe 4468 Calibration Procedure...