PXIe-4468 Calibration Procedure
© National Instruments Corporation
27
Initial Test Connection
Note
Initial test connection is same as AO Differential Offset, as shown in
or
Verification Procedure
Complete the following procedure to verify the AO gain flatness.
Repeat 2 times, once for each channel.
1.
Connect the AO channel under test to the DMM as shown in
Repeat 2 times, once for each output terminal configuration.
Repeat 4 times, once for each AO Attenuation.
Repeat 3 times, once for each frequency.
2.
Configure the DMM using the information in
Table 16
.
3.
Configure an AO Voltage channel task using the values in
Table
21
.
Table 21:
AO Setup
Configuration
Value
Physical Channels
Dev_name
/ao
x
AO Gain*
Gain from
Output Terminal Configuration
Configuration from
Rate
200000 S/s
Sample Mode
Continuous Sample
Samples Per Channel
10000