PXIe-4468 Calibration Procedure
© National Instruments Corporation
29
AO Common-Mode Gain and Offset Verification
Test Limits
Table 23:
AO Common-Mode Gain and Offset Limits
Test Point (V)
As-Found Test Limit
As-Left Test Limit
Lower Limit
(V)
Upper Limit
(V)
Lower Limit
(V)
Upper Limit
(V)
0
-0.00600
0.00600
-0.00152
0.00152
1
0.99325
1.00675
0.99790
1.00210
2
1.99250
2.00750
1.99732
2.00268
3
2.99175
3.00825
2.99674
3.00326
4
3.99100
4.00900
3.99616
4.00384
5
4.99025
5.00975
4.99558
5.00442
Initial Test Connection
Figure 7.
Initial Connection for AO Common-Mode Gain and Offset—BNC
1. DMM
4. BNC – BNC Cable
2. Banana – BNC Adapter
5. PXIe-4468 with BNC
3 Shorting Cable (Banana – Banana)