PXIe-4468 Calibration Procedure
© National Instruments Corporation
15
20
0.99931 * V
1kHz
1.00069 * V
1kHz
45
0.99713 * V
1kHz
1.00288 * V
1kHz
100
0.99083 * V
1kHz
1.00925 * V
1kHz
Initial Test Connection
Note
Initial test connection is same as AI Offset, as shown in
Verification Procedure
Complete the following procedure to verify the AI flatness accuracy.
Repeat 2 times, once for each channel.
1.
Connect the AI channel under test to the Calibrator as shown in
or
Repeat 6 times, once for each gain setting.
Repeat 4 times, once for each frequency in Table 6.
2.
Configure the calibrator to generate an output with the frequency
listed in
Table 7
.
Table 7:
AI Flatness Inputs
Device Gain (dB)
Configuration
Value
Amplitude (Vrms)
Frequency (Hz)
-20
6.3
1000
-10
6.3
1000
0
6.3
1000