PXIe-4468 Calibration Procedure
34
ni.com/services
Excitation (IEPE)
0 (A)
Acquisition mode
Finite number of samples
Rate
40000.0
Samples per channel
2560000
4.
Start the task.
5.
Acquire reading with the DUT.
6.
Measure the frequency using the Extract Single Tone Information VI of
the acquired signal around 10 kHz. Compare the measured frequency
to the limits in
7.
Stop and clear the task. Stop generating a signal with the function
generator.
Perform Adjustment
AI/AO Gain and Offset Adjustment
Perform an adjustment at least once within the calibration interval.
Adjustment automatically updates the calibration constants, the date, and
the temperature in the DUT EEPROM. If the DUT passes the verification
procedures within the As-Left test limits, an adjustment is not required.
Proceed to the
Update the Onboard Calibration Information
section.
Note
Initial test connection is same as AI Offset, as shown in
Adjustment Procedure
1.
Initialization