NI PXIe-4340 Calibration Procedure

|

© National Instruments

|

7

3.

Refer to Table 4 to configure the PXI-4070 and acquire a voltage measurement for AC 
Amplitude.

Table 3.  

PXI-4070 Voltage Measurement Setup for DC Offset

Configuration

Value

Function

DC Volts

Range

10 V

Digital Resolution

6.5 digits

Aperture Time

100 ms

Autozero

Once

ADC Calibration

On

Input Impedance

10 MΩ

DC Noise Rejection

High Order

Number of Averages

10

Power Line Frequency

Dependent upon the local power line characteristics.

Table 4.  

PXI-4070 Voltage Measurement Setup for AC Amplitude

Configuration

Value

Function

AC Volts, DC Coupled

Range

50 V

Digital Resolution

6.5 digits

Aperture Time

200 ms

Autozero

Once

ADC Calibration

On

Input Impedance

1 MΩ

Minimum Frequency

20

Number of Averages

10

Power Line Frequency

Dependent upon the local power line characteristics.

Содержание PXIe-4340

Страница 1: ...rldwide Support and Services 10 Software Calibrating the NI PXIe 4340 requires the installation of NI DAQmx on the calibration system Driver support for calibrating the PXIe 4340 was first available i...

Страница 2: ...nformation NI PXIe 4340 Device Specifications PXIe 4340 specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx NI DAQmx Help Information...

Страница 3: ...not available select a substitute using the requirements listed in Table 1 Table 1 Recommended Equipment for PXIe 4340 Verification Equipment Recommended Model Requirements DMM PXI 4070 Use a DMM that...

Страница 4: ...he TB 4340 Figure 1 TB 4340 Circuit Board Parts Locator Diagram C1 B1 A1 COPYRIGHT 2015 S N 159026B 01L NI TB 4340 EX EX RS RS AI AI CT AIGND CHANNEL 0 EXCITATION VERIFICATION EX EX RS RS AI AI CT AIG...

Страница 5: ...minate noise and thermal offsets Maintain an ambient temperature of 23 C 5 C The PXIe 4340 temperature will be greater than the ambient temperature Maintain an internal device temperature range of Tca...

Страница 6: ...the device name as it appears in MAX Accuracy Verification The following performance verification procedures describe the sequence of operations and provides the test points required to verify the PXI...

Страница 7: ...zero Once ADC Calibration On Input Impedance 10 M DC Noise Rejection High Order Number of Averages 10 Power Line Frequency Dependent upon the local power line characteristics Table 4 PXI 4070 Voltage...

Страница 8: ...hese values the devices passes the DC Offset test at this Excitation Voltage Table 5 PXIe 4340 AI LVDT Measurement Setup Configuration Value Channel Name Dev1 aix where x refers to the channel number...

Страница 9: ...8 Leave all other pins disconnected 2 Configure the channel as shown in Table 5 for 7 Vrms excitation 3 Perform the configured AI LVDT measurement and average the acquired measurements 4 Compare the r...

Страница 10: ...ompleted initialize and close a calibration sessions for Dev1 using the DAQmx Initialize External Calibration VI and the DAQmx Close External Calibration VI Specifications Refer to the NI PXIe 4340 De...

Отзывы: