background image

NI PXIe-4340 Calibration Procedure

|

© National Instruments

|

9

g.

Compare the AC Amplitude measurement to the Excitation Accuracy Limits in 
Table 7. If the measurement is between these values, the device passes the 
AC Amplitude test at this Excitation Voltage.

5.

Repeat step 4 for the other Voltage Excitation Values in Table 7.

6.

Repeat steps 1 to 5 for each additional channel to be verified.

7.

Disconnect the PXI-4070 from the TB-4340.

Ratiometric Accuracy

Complete the following steps to connect the TB-4340 for ratiometric accuracy verification:
1.

Install the PXIe-4340 and the TB-4340 in the PXI Express chassis according to the 
instructions in the 

NI PXIe-4340 and TB-4340 User Guide and Terminal Block 

Specifications

.

2.

Connect AI+ and AI- according to the measurements in Table 8 for each measurement.

Complete the following steps for each measurement test point to verify the ratiometric accuracy 
of the PXIe-4340:
1.

Connect the channel as shown in Table 8. Leave all other pins disconnected.

2.

Configure the channel as shown in Table 5 for 7 Vrms excitation.

3.

Perform the configured AI LVDT measurement and average the acquired measurements.

4.

Compare the resulting average to the 

Minimum Limit

 and 

Maximum Limit

 values shown in 

Table 8. If the result is between these values, the device passes the test.

Table 7.  

Excitation Accuracy Limits

Measurement

Nominal

Maximum Limit

Minimum Limit

Excitation Amplitude 7 Vrms Point

7 Vrms

7.35 Vrms

6.65 Vrms

Excitation Amplitude 5 Vrms Point

5 Vrms

5.25 Vrms

4.75 Vrms

Excitation Amplitude 3 Vrms Point

3 Vrms

3.15 Vrms

2.85 Vrms

Excitation Amplitude 1 Vrms Point

1 Vrms

1.05 Vrms

0.95 Vrms

Table 8.  

Verification Connections

Measurement

Short AI+ to

Short AI- to

Nominal 
Reading

Maximum 

Limit

Minimum 

Limit

1 V/V

EX+

EX-

1 V/V

1.0004000 V/V

0.999600 V/V

0 V/V

AI-

AI+

0 V/V

0.000150 V/V

-0.000150 V/V

-1 V/V

EX-

EX+

-1 V/V

-0.999600 V/V

-1.000400 V/V

Note

: NI-DAQmx returns units scaled for length, while the PXIe-4340 accuracy is specified as a voltage 

ratio. With sensitivity set to 1 mV/V/mm and output units set to meters as in Table 5, 1 meter is numerically 
equivalent to 1 V/V.

Содержание PXIe-4340

Страница 1: ...rldwide Support and Services 10 Software Calibrating the NI PXIe 4340 requires the installation of NI DAQmx on the calibration system Driver support for calibrating the PXIe 4340 was first available i...

Страница 2: ...nformation NI PXIe 4340 Device Specifications PXIe 4340 specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx NI DAQmx Help Information...

Страница 3: ...not available select a substitute using the requirements listed in Table 1 Table 1 Recommended Equipment for PXIe 4340 Verification Equipment Recommended Model Requirements DMM PXI 4070 Use a DMM that...

Страница 4: ...he TB 4340 Figure 1 TB 4340 Circuit Board Parts Locator Diagram C1 B1 A1 COPYRIGHT 2015 S N 159026B 01L NI TB 4340 EX EX RS RS AI AI CT AIGND CHANNEL 0 EXCITATION VERIFICATION EX EX RS RS AI AI CT AIG...

Страница 5: ...minate noise and thermal offsets Maintain an ambient temperature of 23 C 5 C The PXIe 4340 temperature will be greater than the ambient temperature Maintain an internal device temperature range of Tca...

Страница 6: ...the device name as it appears in MAX Accuracy Verification The following performance verification procedures describe the sequence of operations and provides the test points required to verify the PXI...

Страница 7: ...zero Once ADC Calibration On Input Impedance 10 M DC Noise Rejection High Order Number of Averages 10 Power Line Frequency Dependent upon the local power line characteristics Table 4 PXI 4070 Voltage...

Страница 8: ...hese values the devices passes the DC Offset test at this Excitation Voltage Table 5 PXIe 4340 AI LVDT Measurement Setup Configuration Value Channel Name Dev1 aix where x refers to the channel number...

Страница 9: ...8 Leave all other pins disconnected 2 Configure the channel as shown in Table 5 for 7 Vrms excitation 3 Perform the configured AI LVDT measurement and average the acquired measurements 4 Compare the r...

Страница 10: ...ompleted initialize and close a calibration sessions for Dev1 using the DAQmx Initialize External Calibration VI and the DAQmx Close External Calibration VI Specifications Refer to the NI PXIe 4340 De...

Отзывы: