CALIBRATION PROCEDURE

NI PXIe-4340

4 Ch, 24-bit, 25.6 kS/s Simultaneous AC LVDT Input Module

This document contains the verification procedure for the National Instruments PXIe-4340 
module.

Contents

Software.................................................................................................................................... 1
Documentation.......................................................................................................................... 2
PXIe-4340 Verification ............................................................................................................ 3

Test Equipment................................................................................................................. 3

Connecting the TB-4340 .......................................................................................... 4

Test Conditions................................................................................................................. 5
Initial Setup....................................................................................................................... 6
Accuracy Verification....................................................................................................... 6

Excitation Accuracy ................................................................................................. 6
Ratiometric Accuracy ............................................................................................... 9

EEPROM Update ............................................................................................................. 10

Specifications............................................................................................................................ 10
Worldwide Support and Services ............................................................................................. 10

Software

Calibrating the NI PXIe-4340 requires the installation of NI-DAQmx on the calibration system. 
Driver support for calibrating the PXIe-4340 was first available in NI-DAQmx 16.1. For the list 
of devices supported by a specific release, refer to the 

NI-DAQmx Readme

, available on the 

version-specific download page or installation media.

You can download NI-DAQmx from 

ni.com/downloads

. NI-DAQmx supports LabVIEW, 

LabWindows

/CVI

, C/C++, C#, and Visual Basic .NET. When you install NI-DAQmx, you 

only need to install support for the application software that you intend to use.

ni.com/manuals

Deutsch

Français

Содержание PXIe-4340

Страница 1: ...rldwide Support and Services 10 Software Calibrating the NI PXIe 4340 requires the installation of NI DAQmx on the calibration system Driver support for calibrating the PXIe 4340 was first available i...

Страница 2: ...nformation NI PXIe 4340 Device Specifications PXIe 4340 specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx NI DAQmx Help Information...

Страница 3: ...not available select a substitute using the requirements listed in Table 1 Table 1 Recommended Equipment for PXIe 4340 Verification Equipment Recommended Model Requirements DMM PXI 4070 Use a DMM that...

Страница 4: ...he TB 4340 Figure 1 TB 4340 Circuit Board Parts Locator Diagram C1 B1 A1 COPYRIGHT 2015 S N 159026B 01L NI TB 4340 EX EX RS RS AI AI CT AIGND CHANNEL 0 EXCITATION VERIFICATION EX EX RS RS AI AI CT AIG...

Страница 5: ...minate noise and thermal offsets Maintain an ambient temperature of 23 C 5 C The PXIe 4340 temperature will be greater than the ambient temperature Maintain an internal device temperature range of Tca...

Страница 6: ...the device name as it appears in MAX Accuracy Verification The following performance verification procedures describe the sequence of operations and provides the test points required to verify the PXI...

Страница 7: ...zero Once ADC Calibration On Input Impedance 10 M DC Noise Rejection High Order Number of Averages 10 Power Line Frequency Dependent upon the local power line characteristics Table 4 PXI 4070 Voltage...

Страница 8: ...hese values the devices passes the DC Offset test at this Excitation Voltage Table 5 PXIe 4340 AI LVDT Measurement Setup Configuration Value Channel Name Dev1 aix where x refers to the channel number...

Страница 9: ...8 Leave all other pins disconnected 2 Configure the channel as shown in Table 5 for 7 Vrms excitation 3 Perform the configured AI LVDT measurement and average the acquired measurements 4 Compare the r...

Страница 10: ...ompleted initialize and close a calibration sessions for Dev1 using the DAQmx Initialize External Calibration VI and the DAQmx Close External Calibration VI Specifications Refer to the NI PXIe 4340 De...