PXIe-4310 and TB-4310 (10V)/TB-4310 (600V)
|
© National Instruments
|
3
Caution
To ensure the specified EMC performance, operate this product only with
shielded cables and shielded accessories. Use only twisted, shielded pair cables
(Belden 83319 or equivalent) for channel connections. Refer to the
PXIe-4310 User
Manual
for specific guidance on terminating cable shields.
Safety Guidelines for Hazardous Voltages
If hazardous voltages are connected to the module, take the following precautions. A hazardous
voltage is a voltage greater than 42.4 V
pk
or 60 VDC to earth ground.
Caution
Ensure that hazardous voltage wiring is performed only by qualified
personnel adhering to local electrical standards.
Caution
You must install mating connectors according to local safety codes and
standards and according to the specifications provided by the manufacturer. You are
responsible for verifying the safety compliance of third-party connectors and their
usage according to the relevant standard(s), including UL and CSA in North America
and IEC and VDE in Europe.
Caution
Make sure that devices and circuits connected to the module are properly
insulated from human contact.
Caution
When module terminals are hazardous voltage
LIVE (>42.4 V
pk
/60 VDC), you must ensure that devices and circuits connected to
the module are properly insulated from human contact.
Caution
Do not mix hazardous voltage circuits and human-accessible circuits on
the same module.
What You Need to Get Started
To set up and use a PXIe-4310 module with a TB-4310 (10V)/TB-4310 (600V) terminal block,
you need the following items:
Hardware
–
PXIe-4310 module
–
TB-4310 (10V)/TB-4310 (600V) terminal block
–
PXI Express chassis
–
Cabling and sensors as required for your application
Tools
–
Number 1 and 2 Phillips-head screwdrivers
–
1/8 in. flathead screwdriver
–
Long-nose pliers