6
|
ni.com
|
PXIe-4310 and TB-4310 (10V)/TB-4310 (600V)
8.
Secure the module to the chassis using the front-panel captive screws, shown in Figure 2.
Tighten the screws to 0.31 N · m (2.7 lb · in.).
9.
Install a filler panel in any empty slots.
Figure 2.
Installing PXIe-4310
Modules
Connect the Signals
Caution
To ensure the specified EMC performance, operate this product only with
shielded cables and shielded accessories. Use only twisted, shielded pair cables for
channel connections. Refer to the
PXIe-4310 User Manual
for specific guidance on
terminating cable shields.
Caution
Do not supply hazardous voltages (>30 V
rms
/42.4 V
pk
/60 VDC) to the
terminal block without the terminal block being connected to the PXIe-4310.
A terminal block is required to make connections to the module. To connect signals to the
TB-4310 (10V)/TB-4310 (600V) refer to Figures 3 and 4 while completing the following steps:
Note
You can find the pinout names and locations in Measurement & Automation
Explorer (MAX) at any time by right-clicking the device name under
Devices and
Interfaces
and selecting
Device Pinouts
.
1.
Loosen the captive top cover screws and remove the top cover.
2.
Loosen the strain-relief screws and remove the strain-relief bar.
3.
Prepare the shielded signal wire by stripping the insulation no more than 8 mm (0.24 in.).
4.
Run the shielded signal wires through the strain-relief opening.