NI DCM-23XX Getting Started Guide
|
© National Instruments
|
11
1.
Open NI Software Calibration Management (SCM) Console on the host PC.
2.
Select
File
»
Select Target
.
3.
Choose your DCM on the
Select Target
drop-down.
4.
Select
5.
Click
OK
.
6.
Click
Start Host Application
on the SCM Console.
Note
The DCM is shipped supporting the latest version of SCM. If you have
trouble launching the DCM DSI with the latest available version of SCM on
the software on your DCM may need to be updated. Refer to the
NI DCM-23XX DSI
User’s Guide
for more information.
Troubleshooting the DCM
Safe Mode and IP Reset
The following figure shows the reset behavior of the DCM.
Figure 3.
Reset Behavior
LED Indicators
The following table describes the DCM LED indicators.
Run Mode
• Network Settings Reset
• Console Out Enabled
• RT Startup App Disabled
• FPGA Startup App Disabled
Safe Mode with
IP Reset
• Console Out Enabled
• RT Startup App Disabled
• FPGA Startup App Disabled
Safe Mode
Cycle Power
Cycle Power
+AI11 (pin 63) to Batt+
+AI12 (pin 55) to Batt+
Cycle Power
+AI12 (pin 55) to Batt+
Cycle Power