National Instruments Corporation
27
NI 4070/4072 Calibration Procedure
24. Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function
=
NIDMM_VAL_2_WIRE_RES
•
Range
=
1e6
•
Resolution
=
1
25. Call
niDMM_Read
. Verify that the result falls between the tolerances
listed in Table 20.
26. Output 100 k
Ω
on the calibrator without external sense or 2-wire
compensation.
27. Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function
=
NIDMM_VAL_2_WIRE_RES
•
Range
=
100e3
•
Resolution
=
0.1
28. Call
niDMM_Read
. Verify that the result falls between the tolerances
listed in Table 20.
29. Connect the NI 4070/4072 and the Fluke 5700A/5720A calibrator
using two sets of Fluke 5440 cables, a double banana plug, and two
insulated, low EMF spade lugs, as shown in Figure 3. Table 10 lists the
cable connections.