National Instruments Corporation
31
NI 4070/4072 Calibration Procedure
5.
Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function
=
NIDMM_VAL_DC_CURRENT
•
Range
=
0.02
•
Resolution
=
20e–9
6.
Call
niDMM_Read
. Verify that this measurement falls between the
limits listed in Table 21.
7.
Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function
=
NIDMM_VAL_DC_CURRENT
•
Range
=
0.2
•
Resolution
=
200e–9
8.
Call
niDMM_Read
. Verify that this measurement falls between the
limits listed in Table 21.
9.
Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function
=
NIDMM_VAL_DC_CURRENT
•
Range
=
1
•
Resolution
=
1e–6
10. Call
niDMM_Read
. Verify that this measurement falls between the
limits listed in Table 21.
11. Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function
=
NIDMM_VAL_DC_CURRENT
•
Range
=
0.02
•
Resolution
=
20e–9
12. Call
niDMM_Read
to configure the NI 4070/4072 for a current mode
before applying current.
13. Output 20 mA on the calibrator.
14. Call
niDMM_Read
. Verify that this measurement falls between the
limits listed in Table 21.
15. Output –20 mA on the calibrator.
16. Call
niDMM_Read
. Verify that this measurement falls between the
limits listed in Table 21.
17. Output 200 mA on the calibrator.