NI 4070/4072 Calibration Procedure
26
ni.com
10. Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function
=
NIDMM_VAL_2_WIRE_RES
•
Range
=
100
•
Resolution
=
100e–6
11. Call
niDMM_ConfigureOffsetCompOhms
with
OffsetCompOhms
set to
NIDMM_VAL_OFFSET_COMP_OHMS_ON
.
12. Call
niDMM_Read
. Verify that this measurement falls between the
limits listed in Table 20.
13. Remove the shorting bar from the NI 4070/4072.
14. Reset the calibrator.
15. Fasten the connectors on one end of the Fluke 5440 cable to the
NI 4070/4072, and fasten the connectors on the other end of the cable
to the appropriate calibrator binding posts. Figure 1 shows the correct
connections. Table 2 lists the cable connections.
16. Wait two minutes for the thermal EMF to stabilize if the Fluke 5440
cable was not previously used in this configuration.
17. Output 100 M
Ω
on the calibrator without external sense or 2-wire
compensation.
18. Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function
=
NIDMM_VAL_2_WIRE_RES
•
Range
=
100e6
•
Resolution
=
100
19. Call
niDMM_Read
. Verify that the result falls between the tolerances
listed in Table 20.
20. Output 10 M
Ω
on the calibrator without external sense or 2-wire
compensation.
21. Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function
=
NIDMM_VAL_2_WIRE_RES
•
Range
=
10e6
•
Resolution
=
10
22. Call
niDMM_Read
. Verify that the result falls between the tolerances
listed in Table 20.
23. Output 1 M
Ω
on the calibrator without external sense or 2-wire
compensation.