background image

NI 4070/4072 Calibration Procedure

24

ni.com

5.

Refer to Table 9 for the appropriate calibrator output and function 
parameter values as you complete the following steps:

a.

On the calibrator, output the value listed in the 

Calibrator Output

 

column in Table 9 for the current iteration. Make sure that the 
external sense is turned on but 2-wire compensation is turned off.

Note

After setting the calibrator output to 0

Ω

 in the seventh iteration, you do 

not

 need to 

continually set the calibrator to 0

Ω

 for iterations 8 through 12.

b.

Call 

niDMM_ConfigureMeasurement

 with the parameters set 

as shown in Table 9 for the current iteration.

c.

Call 

niDMM_ConfigureOffsetCompOhms

 with 

OffsetCompOhms

 set to either 

NIDMM_VAL_OFFSET_COMP_OHMS_ON

 or 

NIDMM_VAL_OFFSET_COMP_OHMS_OFF

 according to Table 9 for 

the current iteration.

d.

Call 

niDMM_Read

. Verify that this measurement falls between the 

tolerances listed in Table 19. Tolerances are provided instead of 
absolute limits because your calibrator will have different discrete 
resistance values.

6.

Repeat step 5 for each of the remaining iterations listed in Table 9.

Table 9.  

niDMM_ConfigureMeasurement Parameters

Iteration

Calibrator

Output

niDMM_ConfigureMeasurement Parameters

OffsetCompOhms

Function

Range

Resolution

1

10 M

Ω

NIDMM_VAL_4_WIRE_RES

10e6

10

OFF

2

1 M

Ω

NIDMM_VAL_4_WIRE_RES

1e6

1

OFF

3

100 k

Ω

NIDMM_VAL_4_WIRE_RES

100e3

0.1

OFF

4

10 k

Ω

NIDMM_VAL_4_WIRE_RES

10e3

0.01

ON

5

1 k

Ω

NIDMM_VAL_4_WIRE_RES

1e3

1e–3

ON

6

100

Ω

NIDMM_VAL_4_WIRE_RES

100

100e–6

ON

7

0

Ω

NIDMM_VAL_4_WIRE_RES

10e6

10

OFF

8

0

Ω

NIDMM_VAL_4_WIRE_RES

1e6

1

OFF

9

0

Ω

NIDMM_VAL_4_WIRE_RES

100e3

0.1

OFF

10

0

Ω

NIDMM_VAL_4_WIRE_RES

10e3

0.01

ON

11

0

Ω

NIDMM_VAL_4_WIRE_RES

1e3

1e–3

ON

12

0

Ω

NIDMM_VAL_4_WIRE_RES

100

100e–6

ON

Содержание FlexDMM PCI-4070

Страница 1: ...tion on calibration visit ni com calibration Contents Conventions 2 Software Requirements 2 Documentation Requirements 3 Calibration Function Reference 3 Password 3 Calibration Interval 3 Test Equipment 4 Test Conditions 5 Calibration Procedures 6 Initial Setup 6 Verification Procedures 7 Adjustment Procedures 39 Verification Limits 58 DC Voltage 58 AC Voltage 59 4 Wire Resistance 60 2 Wire Resist...

Страница 2: ... names italic Italic text denotes variables emphasis a cross reference hardware labels or an introduction to a key concept Italic text also denotes text that is a placeholder for a word or value that you must supply monospace Text in this font denotes text or characters that you should enter from the keyboard sections of code programming examples and syntax examples This font is also used for the ...

Страница 3: ...quency verification procedure TB 2715 Terminal Block Installation Guide Note to Users About Your 6608 Device Calibration Function Reference For detailed information about the NI DMM calibration functions used in this procedure refer to the LabVIEW reference or the C CVI VB reference sections of the NI Digital Multimeters Help located at Start All Programs National Instruments NI DMM Documentation ...

Страница 4: ...LO input banana plug connectors on the NI 4071 Pomona 5145 insulated double banana plug shorting bar Two sets of banana to banana cables with length not to exceed 4 in Two Pomona B 4 banana to banana patch cords cables Double banana plug with binding posts Pomona 5405 Binding Post Insulated low thermal electromotive force EMF spade lugs Two Pomona 2305 lugs Chassis National Instruments PXI chassis...

Страница 5: ...rnack PE3005 banana to banana coaxial cables with length 4 inches and total capacitance 40 pF Before performing the verification procedure you should know the total capacitance up to the end of the banana connectors that plug into the NI 4072 Test Conditions Follow these guidelines to optimize the connections and the environment during calibration Ensure that the PXI chassis fan speed is set to HI...

Страница 6: ...flow for verification 3 Adjustment Procedures Perform an external adjustment of the device that adjusts the calibration constants with respect to standards of known values Figure 6 shows the procedural flow for adjustment 4 Reverification Repeat the verification procedure to ensure that the device is operating within its specifications after adjustment These steps are described in more detail in t...

Страница 7: ... the entire Verifying AC Current section if necessary The parameters Range Resolution and Sample Interval used in function calls throughout this section have floating point values For example if Range 1 the floating point value is 1 0 The parameters Trigger Count Sample Count Array Size and ParamValue have integer values Refer to the NI Digital Multimeters Help for more information about parameter...

Страница 8: ...e_Value NIDMM_VAL_10_MEGAOHM 8 Call niDMM_Read Verify that this measurement falls between the limits listed in Table 17 9 Call niDMM_ConfigureMeasurement with the following parameters Function NIDMM_VAL_DC_VOLTS Range 10 Resolution 10e 6 10 Set the input resistance of the NI 4070 4072 to 10 GΩ by calling niDMM_SetAttributeViReal64 with the following parameters Attribute_ID NIDMM_ATTR_INPUT_RESISTA...

Страница 9: ...owing parameters Function NIDMM_VAL_DC_VOLTS Range 300 Resolution 300e 6 18 Set the input resistance of the NI 4070 4072 to 10 MΩ by calling niDMM_SetAttributeViReal64 with the following parameters Attribute_ID NIDMM_ATTR_INPUT_RESISTANCE Attribute_Value NIDMM_VAL_10_MEGAOHM 19 Call niDMM_Read Verify that this measurement falls between the limits listed in Table 17 20 Remove the shorting bar from ...

Страница 10: ...ttributeViReal64 with the following parameters Attribute_ID NIDMM_ATTR_INPUT_RESISTANCE Attribute_Value NIDMM_VAL_GREATER_THAN_10_GIGAOHM 1 NI 4070 4072 2 Fluke 5700A 5720A Calibrator 3 Fluke 5440 Cable Table 2 Fluke 5440 Cable Connections Banana Plug Connector NI 4070 4072 Banana Plug Color Fluke 5440 Cable Binding Post Label Fluke 5700A 5720A Calibrator HI Red OUTPUT HI LO Black OUTPUT LO No con...

Страница 11: ...ttribute_ID NIDMM_ATTR_INPUT_RESISTANCE Attribute_Value NIDMM_VAL_10_MEGAOHM 30 Call niDMM_ConfigureMultiPoint with the following parameters Trigger Count 1 Sample Count 10 Sample Trigger NIDMM_VAL_IMMEDIATE SampleInterval 1 31 Call niDMM_ReadMultiPoint with the following parameters Maximum Time NIDMM_VAL_TIME_LIMIT_AUTO Array Size 10 Average the results by summing the returned reading array of th...

Страница 12: ...cked to 2 2 V This range prevents a 50 Ω calibrator output resistance from creating a voltage divider with the internal resistance of the NI 4070 4072 39 Set the input resistance of the NI 4070 4072 to 10 GΩ by calling niDMM_SetAttributeViReal64 with the following parameters Attribute_ID NIDMM_ATTR_INPUT_RESISTANCE Attribute_Value NIDMM_VAL_GREATER_THAN_10_GIGAOHM 40 Call niDMM_Read Subtract the p...

Страница 13: ...AttributeViReal64 with the following parameters Attribute_ID NIDMM_ATTR_INPUT_RESISTANCE Attribute_Value NIDMM_VAL_GREATER_THAN_10_GIGAOHM 51 Call niDMM_Read Verify that this measurement falls between the limits listed in Table 17 52 Set the input resistance of the NI 4070 4072 to 10 MΩ by calling niDMM_SetAttributeViReal64 with the following parameters Attribute_ID NIDMM_ATTR_INPUT_RESISTANCE Att...

Страница 14: ...070 4072 to 10 MΩ by calling niDMM_SetAttributeViReal64 with the following parameters Attribute_ID NIDMM_ATTR_INPUT_RESISTANCE Attribute_Value NIDMM_VAL_10_MEGAOHM 64 Call niDMM_Read Verify that this measurement falls between the limits listed in Table 17 65 Output 100 V on the calibrator Caution Avoid touching the connections when generating a high voltage from the calibrator 66 Call niDMM_Config...

Страница 15: ...ant to continue verifying other modes go to the Verifying AC Voltage section If you do not want to verify other modes and you are performing a post adjustment verification go to the Completing the Adjustment Procedures section If you do not want to verify any additional modes and you are performing a pre adjustment verification call niDMM_close to close the session Verifying AC Voltage To verify A...

Страница 16: ...ureMeasurement with the following parameters Function NIDMM_VAL_AC_VOLTS_DCCOUPLED Range 0 05 Resolution 50e 9 12 Call niDMM_Read Verify that this measurement falls between the limits listed in Table 18 13 Refer to Table 3 for the appropriate calibrator outputs and parameter values as you complete the following steps a On the calibrator output the value listed in the Calibrator Output column in Ta...

Страница 17: ...d the remaining parameters as shown in Table 4 for the current iteration Table 3 niDMM_ConfigureMeasurement Parameters Iteration Calibrator Output niDMM_ConfigureMeasurement Parameters Amplitude Frequency Function Range Resolution 1 50 mV 50 Hz NIDMM_VAL_AC_VOLTS 0 05 50e 9 50 mV 50 Hz NIDMM_VAL_AC_VOLTS_DCCOUPLED 0 05 50e 9 2 50 mV 1 kHz NIDMM_VAL_AC_VOLTS 0 05 50e 9 50 mV 1 kHz NIDMM_VAL_AC_VOLT...

Страница 18: ...arameters Iteration Calibrator Output niDMM_ConfigureMeasurement Parameters Amplitude Frequency Function Range Resolution 1 500 mV 50 Hz NIDMM_VAL_AC_VOLTS 0 5 500e 9 500 mV 50 Hz NIDMM_VAL_AC_VOLTS_DCCOUPLED 0 5 500e 9 2 500 mV 1 kHz NIDMM_VAL_AC_VOLTS 0 5 500e 9 500 mV 1 kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 0 5 500e 9 3 500 mV 1 kHz NIDMM_VAL_AC_VOLTS 5 5e 6 500 mV 1 kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED...

Страница 19: ... Table 18 23 Output 50 V at 30 Hz on the calibrator Table 5 niDMM_ConfigureMeasurement Parameters Iteration Calibrator Output niDMM_ConfigureMeasurement Parameters Amplitude Frequency Function Range Resolution 1 5 V 50 Hz NIDMM_VAL_AC_VOLTS 5 5e 6 5 V 50 Hz NIDMM_VAL_AC_VOLTS_DCCOUPLED 5 5e 6 2 5 V 1 kHz NIDMM_VAL_AC_VOLTS 5 5e 6 5 V 1 kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 5 5e 6 3 5 V 1 kHz NIDMM_VAL_...

Страница 20: ...its listed in Table 18 d Call niDMM_ConfigureMeasurement again changing Mode to NIDMM_VAL_AC_VOLTS_DCCOUPLED e Call niDMM_Read Verify that this measurement falls between the limits listed in Table 18 Table 6 niDMM_ConfigureMeasurement Parameters Iteration Calibrator Output niDMM_ConfigureMeasurement Parameters Amplitude Frequency Function Range Resolution 1 50 V 50 Hz NIDMM_VAL_AC_VOLTS 50 50e 6 5...

Страница 21: ...DMM_VAL_AC_VOLTS and the remaining parameters as shown in Table 7 for the current iteration c Call niDMM_Read Verify that this measurement falls between the limits listed in Table 18 d Call niDMM_ConfigureMeasurement again changing Mode to NIDMM_VAL_AC_VOLTS_DCCOUPLED e Call niDMM_Read Verify that this measurement falls between the limits listed in Table 18 Table 7 niDMM_ConfigureMeasurement Param...

Страница 22: ...re Resistance To verify the 4 wire resistance of the NI 4070 4072 complete the following steps 1 Reset the calibrator 2 Fasten the connectors on one end of each Fluke 5440 cable to the appropriate banana plug connectors on the NI 4070 4072 Fasten the connectors on the other end of each Fluke 5440 cable to the appropriate calibrator binding posts Figure 2 shows the Fluke 5440 cables Table 8 lists t...

Страница 23: ...440 Cables Table 8 Fluke 5440 Cable Connections Fluke 5440 Cable Identification Banana Plug Connector NI 4070 4072 Banana Plug Color Fluke 5440 Cable Binding Post Fluke 5700A 5720A Calibrator First cable HI Red OUTPUT HI LO Black OUTPUT LO No connection Blue V GUARD Second cable HI SENSE Red SENSE HI LO SENSE Black SENSE LO No connection Blue V GUARD HI LO 300V MAX HI LO AUX I O 1A 250V MAX 300V M...

Страница 24: ...ding to Table 9 for the current iteration d Call niDMM_Read Verify that this measurement falls between the tolerances listed in Table 19 Tolerances are provided instead of absolute limits because your calibrator will have different discrete resistance values 6 Repeat step 5 for each of the remaining iterations listed in Table 9 Table 9 niDMM_ConfigureMeasurement Parameters Iteration Calibrator Out...

Страница 25: ... 4070 4072 complete the following steps 1 Plug in the insulated banana plug shorting bar across the HI and LO banana plug connectors on the NI 4070 4072 2 Wait one minute for the thermal EMF to stabilize 3 Call niDMM_reset 4 Call niDMM_ConfigureMeasurement with the following parameters Function NIDMM_VAL_2_WIRE_RES Range 10e3 Resolution 0 01 5 Call niDMM_ConfigureOffsetCompOhms with OffsetCompOhms...

Страница 26: ...t connections Table 2 lists the cable connections 16 Wait two minutes for the thermal EMF to stabilize if the Fluke 5440 cable was not previously used in this configuration 17 Output 100 MΩ on the calibrator without external sense or 2 wire compensation 18 Call niDMM_ConfigureMeasurement with the following parameters Function NIDMM_VAL_2_WIRE_RES Range 100e6 Resolution 100 19 Call niDMM_Read Verif...

Страница 27: ...kΩ on the calibrator without external sense or 2 wire compensation 27 Call niDMM_ConfigureMeasurement with the following parameters Function NIDMM_VAL_2_WIRE_RES Range 100e3 Resolution 0 1 28 Call niDMM_Read Verify that the result falls between the tolerances listed in Table 20 29 Connect the NI 4070 4072 and the Fluke 5700A 5720A calibrator using two sets of Fluke 5440 cables a double banana plug...

Страница 28: ... external sense turned on Table 10 Fluke 5440 Cable Connections Banana Plug Connector NI 4070 4072 Banana Plug Color Fluke 5440 Cable Binding Post Label Fluke 5700A 5720A Calibrator HI Red OUTPUT HI LO Black OUTPUT LO No connection Blue V GUARD HI Red OUTPUT HI SENSE LO Black OUTPUT LO SENSE No connection Blue V GUARD HI LO AUX I O 300V MAX 5V MAX AMPS NI PXI 407x 61 2 Digit FlexDMM 500V MA X CAT ...

Страница 29: ...t the result falls between the tolerances listed in Table 20 39 Output 100 Ω on the calibrator with 2 wire compensation and external sense turned on 40 Call niDMM_ConfigureMeasurement with the following parameters Function NIDMM_VAL_2_WIRE_RES Range 100 Resolution 100e 6 41 Call niDMM_Read Verify that the result falls between the tolerances listed in Table 20 You have completed verifying the 2 wir...

Страница 30: ...able connections Figure 4 Cable Connections for Current 3 Call niDMM_reset to reset the NI 4070 4072 to a known state 4 Set the current output on the calibrator to NORM and output 0 A 1 NI 4070 4072 2 Fluke 5700A 5720A Calibrator 3 Fluke 5440 Cable Table 11 Fluke 5440 Cable Connections Banana Plug Connector NI 4070 4072 Banana Plug Color Fluke 5440 Cable Binding Post Fluke 5700A 5720A Calibrator H...

Страница 31: ... niDMM_ConfigureMeasurement with the following parameters Function NIDMM_VAL_DC_CURRENT Range 1 Resolution 1e 6 10 Call niDMM_Read Verify that this measurement falls between the limits listed in Table 21 11 Call niDMM_ConfigureMeasurement with the following parameters Function NIDMM_VAL_DC_CURRENT Range 0 02 Resolution 20e 9 12 Call niDMM_Read to configure the NI 4070 4072 for a current mode befor...

Страница 32: ...Range 1 Resolution 1e 6 24 Call niDMM_Read Verify that this measurement falls between the limits listed in Table 21 25 Output 1 A on the calibrator 26 Call niDMM_Read Verify that this measurement falls between the limits listed in Table 21 You have completed verifying the DC current of the NI 4070 4072 Select one of the following options If you want to continue verifying other modes go to the Veri...

Страница 33: ...tion 10e 9 5 Call niDMM_ConfigureAutoZeroMode with the following parameters Auto Zero Mode NIDMM_VAL_AUTO_ZERO_ON 6 Call niDMM_Read to configure the NI 4070 4072 for a current mode before applying current 7 Output 1 mA at 1 kHz on the calibrator with the current output set to NORM 8 Call niDMM_Read Verify that this measurement falls between the limits listed in Table 22 9 Output 10 mA at 1 kHz on ...

Страница 34: ...ing the Adjustment Procedures section If you do not want to verify any additional modes and you are performing a pre adjustment verification call niDMM_close to close the session Verifying Frequency Note The frequency of the NI 4070 4072 is not user adjustable If this verification procedure indicates that the frequency is out of specification return the NI 4070 4072 to NI for repair To verify the ...

Страница 35: ...ifying Capacitance and Inductance NI 4072 Only section If you do not want to verify other modes and you are performing a post adjustment verification go to the Completing the Adjustment Procedures section If you do not want to verify any additional modes and you are performing a pre adjustment verification call niDMM_close to close the session Verifying Capacitance and Inductance NI 4072 Only This...

Страница 36: ...on capacitors to the NI 4072 banana plug connectors minimize noise by ensuring that the cables remain fixed and do not move during the measurement Keep direct contact with the verification capacitors to a minimum so that they are constantly kept at the ambient temperature After connecting a capacitor to the NI 4072 terminals NI recommends waiting 30 seconds for the capacitor temperature to stabili...

Страница 37: ...ances listed in Table 24 10 Remove the shorting bar and plug the 270 pF verification capacitor into the HI and LO banana plug connectors of the NI 4072 Remember to wait 30 seconds for the temperature to stabilize before performing the next step 11 Call niDMM_ConfigureMeasurement with the following parameters Function NIDMM_VAL_CAPACITANCE Range 300e 12 Resolution 50e 15 12 Set the number of averag...

Страница 38: ...racy specifications 17 Call niDMM_Read Verify that this measurement falls between the tolerances listed in Table 24 18 Call niDMM_ConfigureMeasurement with the following parameters Function NIDMM_VAL_CAPACITANCE Range 10e 9 Resolution 1e 12 19 Set the number of averages of the NI 4072 to 20 by calling niDMM_SetAttributeViInt32 with the following parameters Attribute_ID NIDMM_ATTR_LC_NUMBER_MEAS_TO...

Страница 39: ...on call niDMM_close to close the session If you are performing a post adjustment verification go to the Completing the Adjustment Procedures section Adjustment Procedures This section explains how to adjust the NI 4070 4072 You can choose to perform these adjustment procedures with or without performing the verification procedures first The parameters Range Resolution Expected Value and Frequency ...

Страница 40: ...tions on calibrating these devices Note Ensure that the calibrator is warmed up for at least 60 minutes before you begin this procedure 3 Reset the calibrator 4 If you have not already done so allow the NI 4070 4072 to warm up for 60 minutes within a powered on PXI chassis or PC Adjusting DC Voltage and Resistance To adjust the DC voltage and resistance of the NI 4070 4072 complete the following s...

Страница 41: ...ll niDMM_CalAdjustGain with the following parameters Mode NIDMM_VAL_DC_VOLTS Range 0 1 Input Resistance NIDMM_VAL_10_MEGAOHM Expected Value 0 1 9 Output 10 V on the calibrator 10 Call niDMM_CalAdjustGain with the following parameters Mode NIDMM_VAL_DC_VOLTS Range 10 Input Resistance NIDMM_VAL_GREATER_THAN_10_GIGAOHM Expected Value 10 11 Output 10 V on the calibrator 12 Call niDMM_CalAdjustGain wit...

Страница 42: ...l sense or 2 wire compensation 22 Call niDMM_CalAdjustGain with the following parameters Mode NIDMM_VAL_2_WIRE_RES Range 10e6 Input Resistance NIDMM_VAL_RESISTANCE_NA Expected Value the display on the calibrator for 0 Ω 23 Output 10 MΩ from the calibrator without external sense 24 Call niDMM_CalAdjustGain with the following parameters Mode NIDMM_VAL_2_WIRE_RES Range 10e6 Input Resistance NIDMM_VAL...

Страница 43: ...ing parameters Mode NIDMM_VAL_2_WIRE_RES Range 100e6 Input Resistance NIDMM_VAL_RESISTANCE_NA Expected Value the display on the calibrator for 100 MΩ 35 Output 0 Ω from the calibrator without external sense or 2 wire compensation 36 Call niDMM_CalAdjustGain with the following parameters Mode NIDMM_VAL_2_WIRE_RES Range 100e6 Input Resistance NIDMM_VAL_RESISTANCE_NA Expected Value the display on the...

Страница 44: ...the calibrator with external sense turned on and 2 wire compensation turned off 45 Call niDMM_CalAdjustGain with the following parameters Mode NIDMM_VAL_4_WIRE_RES Range 100e3 Input Resistance NIDMM_VAL_RESISTANCE_NA Expected Value the display on the calibrator for 100 kΩ 46 Output 0 Ω on the calibrator with external sense turned on and 2 wire compensation turned off 47 Call niDMM_CalAdjustGain wi...

Страница 45: ... compensation turned off 56 Call niDMM_CalAdjustOffset with the following parameters Mode NIDMM_VAL_4_WIRE_RES Range 10e6 Input Resistance NIDMM_VAL_RESISTANCE_NA 57 Call niDMM_CalAdjustOffset with the following parameters Mode NIDMM_VAL_4_WIRE_RES Range 1e6 Input Resistance NIDMM_VAL_RESISTANCE_NA 58 Call niDMM_CalAdjustOffset with the following parameters Mode NIDMM_VAL_4_WIRE_RES Range 1e3 Inpu...

Страница 46: ...DMM_CalAdjustOffset with the following parameters Mode NIDMM_VAL_2_WIRE_RES Range 1e3 Input Resistance NIDMM_VAL_RESISTANCE_NA 68 Call niDMM_CalAdjustOffset with the following parameters Mode NIDMM_VAL_2_WIRE_RES Range 100 Input Resistance NIDMM_VAL_RESISTANCE_NA 69 Call niDMM_CalAdjustMisc with Type set to NIDMM_EXTCAL_MISCCAL_SECTION You have completed adjusting the DC voltage and resistance mod...

Страница 47: ...ust the AC voltage of the NI 4070 4072 complete the following steps 1 Reset the calibrator 2 Fasten the connectors on one end of the Fluke 5440 cable into the appropriate banana plug connectors on the NI 4070 4072 and fasten the connectors on the other end of the cable to the appropriate calibrator binding posts Figure 1 shows the correct connections Table 2 lists the cable connections 3 Refer to ...

Страница 48: ...brator Output niDMM_CalAdjustGain Parameters Amplitude f kHz Mode Range V Input Resistance Expected Value 1 50 mV 1 NIDMM_VAL_AC_VOLTS 0 05 NIDMM_VAL_1_ME GAOHM 0 05 50 mV 1 NIDMM_VAL_AC_VOLTS_ DCCOUPLED 0 05 NIDMM_VAL_1_ME GAOHM 0 05 2 500 mV 1 NIDMM_VAL_AC_VOLTS 0 5 NIDMM_VAL_1_ME GAOHM 0 5 500 mV 1 NIDMM_VAL_AC_VOLTS_ DCCOUPLED 0 5 NIDMM_VAL_1_ME GAOHM 0 5 3 5 V 1 NIDMM_VAL_AC_VOLTS 5 NIDMM_VAL...

Страница 49: ...ter remains the same for all instances of this function c Call niDMM_CalAdjustACFilter again changing Mode to NIDMM_VAL_AC_VOLTS_DCCOUPLED Table 14 niDMM_CalAdjustOffset Parameters Iteration niDMM_CalAdjustOffset Parameters Mode Range V Input Resistance Ω 1 NIDMM_VAL_AC_VOLTS 0 05 NIDMM_VAL_1_MEGAOHM NIDMM_VAL_AC_VOLTS_DCCOUPLED 0 05 NIDMM_VAL_1_MEGAOHM 2 NIDMM_VAL_AC_VOLTS 0 5 NIDMM_VAL_1_MEGAOHM...

Страница 50: ... 50 mV 100 NIDMM_VAL_AC_VOLTS_DCCOUPLED 0 05 100e3 6 50 mV 200 NIDMM_VAL_AC_VOLTS 0 05 200e3 50 mV 200 NIDMM_VAL_AC_VOLTS_DCCOUPLED 0 05 200e3 7 50 mV 300 NIDMM_VAL_AC_VOLTS 0 05 300e3 50 mV 300 NIDMM_VAL_AC_VOLTS_DCCOUPLED 0 05 300e3 8 50 mV 500 NIDMM_VAL_AC_VOLTS 0 05 500e3 50 mV 500 NIDMM_VAL_AC_VOLTS_DCCOUPLED 0 05 500e3 9 500 mV 1 NIDMM_VAL_AC_VOLTS 0 5 1e3 500 mV 1 NIDMM_VAL_AC_VOLTS_DCCOUPL...

Страница 51: ...TS 5 100e3 5 V 100 NIDMM_VAL_AC_VOLTS_DCCOUPLED 5 100e3 22 5 V 200 NIDMM_VAL_AC_VOLTS 5 200e3 5 V 200 NIDMM_VAL_AC_VOLTS_DCCOUPLED 5 200e3 23 5 V 300 NIDMM_VAL_AC_VOLTS 5 300e3 5 V 300 NIDMM_VAL_AC_VOLTS_DCCOUPLED 5 300e3 24 5 V 500 NIDMM_VAL_AC_VOLTS 5 500e3 5 V 500 NIDMM_VAL_AC_VOLTS_DCCOUPLED 5 500e3 25 50 V 1 NIDMM_VAL_AC_VOLTS 50 1e3 50 V 1 NIDMM_VAL_AC_VOLTS_DCCOUPLED 50 1e3 26 50 V 5 NIDMM_...

Страница 52: ... 1e3 34 100 V 5 NIDMM_VAL_AC_VOLTS 300 5e3 100 V 5 NIDMM_VAL_AC_VOLTS_DCCOUPLED 300 5e3 35 100 V 20 NIDMM_VAL_AC_VOLTS 300 20e3 100 V 20 NIDMM_VAL_AC_VOLTS_DCCOUPLED 300 20e3 36 100 V 50 NIDMM_VAL_AC_VOLTS 300 50e3 100 V 50 NIDMM_VAL_AC_VOLTS_DCCOUPLED 300 50e3 37 100 V 100 NIDMM_VAL_AC_VOLTS 300 100e3 100 V 100 NIDMM_VAL_AC_VOLTS_DCCOUPLED 300 100e3 38 100 V 200 NIDMM_VAL_AC_VOLTS 300 200e3 100 V...

Страница 53: ...acy is insignificant for your application you can skip this section and select one of the following options If you skip this section and you are calibrating an NI 4072 go to the Adjusting Capacitance and Inductance NI 4072 Only section If you skip this section and you want to verify the new calibration coefficients before saving them to the EEPROM repeat the Verification Procedures section except ...

Страница 54: ...stance NIDMM_VAL_RESISTANCE_NA Expected Value 0 02 9 Output 0 A on the calibrator 10 Call niDMM_CalAdjustOffset with the following parameters Mode NIDMM_VAL_DC_CURRENT Range 0 02 Input Resistance NIDMM_VAL_RESISTANCE_NA 11 Call niDMM_CalAdjustOffset with the following parameters Mode NIDMM_VAL_AC_CURRENT Range 0 01 Input Resistance NIDMM_VAL_RESISTANCE_NA 12 Output 200 mA on the calibrator 13 Call...

Страница 55: ...MM_VAL_DC_CURRENT Range 1 Input Resistance NIDMM_VAL_RESISTANCE_NA Expected Value 1 21 Output 1 A on the calibrator 22 Call niDMM_CalAdjustGain with the following parameters Mode NIDMM_VAL_DC_CURRENT Range 1 Input Resistance NIDMM_VAL_RESISTANCE_NA Expected Value 1 23 Output 0 A on the calibrator with the current output set to NORM 24 Call niDMM_CalAdjustOffset with the following parameters Mode N...

Страница 56: ...f you do not want to verify the new calibration coefficients go to the Completing the Adjustment Procedures section If you are calibrating an NI 4072 you must complete this section to attain a valid calibration Caution It is necessary to adjust DC voltage and resistance before running these adjustment steps During this procedure be sure to keep hands and any other moving objects away from the fixt...

Страница 57: ...plete the appropriate procedures To finish the calibration and close the session go to the Completing the Adjustment Procedures section Completing the Adjustment Procedures To complete the adjustment procedure for the NI 4070 4072 and close the session call niDMM_CloseExtCal with the following parameter Action NIDMM_EXTCAL_ACTION_SAVE if the results of the calibration were satisfactory and you wan...

Страница 58: ...on the latest specifications DC Voltage Table 17 NI 4070 4072 DC Voltage Verification Limits Calibrator Amplitude Range Input Resistance 2 Year Limits 24 Hour Limits Lower Upper Lower Upper 0 V 1 V 10 GΩ 10 MΩ 6 μV 6 μV 2 μV 2 μV 0 V 10 V 10 GΩ 10 MΩ 60 μV 60 μV 20 μV 20 μV 0 V 100 V 10 MΩ 600 μV 600 μV 200 μV 200 μV 0 V 300 V 10 MΩ 6 mV 6 mV 1 8 mV 1 8 mV 100 mV 100 mV 10 GΩ 10 MΩ 0 099994 V 0 10...

Страница 59: ... 0 04845 V 0 05155 V 500 mV 30 Hz 500 mV DC 0 49945 V 0 50055 V 500 mV 50 Hz 500 mV AC DC 0 49965 V 0 50035 V 500 mV 1 kHz 500 mV AC DC 0 49965 V 0 50035 V 500 mV 1 kHz 5 V AC DC 0 49875 V 0 50125 V 500 mV 20 kHz 500 mV AC DC 0 49965 V 0 50035 V 500 mV 50 kHz 500 mV AC DC 0 49945 V 0 50055 V 500 mV 100 kHz 500 mV AC DC 0 4974 V 0 5026 V 500 mV 300 kHz 500 mV AC DC 0 48475 V 0 51525 V 5 V 30 Hz 5 V...

Страница 60: ... 8305 V 219 1695 V 219 V 20 kHz 300 V AC DC 218 8305 V 219 1695 V 219 V 50 kHz 300 V AC DC 218 7429 V 219 2571 V 219 V 100 kHz 300 V AC DC 217 845 V 220 155 V 70 V 300 kHz 300 V AC DC 67 75 V 72 25 V Table 19 NI 4070 4072 4 Wire Resistance Verification Tolerances Calibrator Resistance Range 2 Year Tolerance ppm of Calibrator Resistance ppm of Range 24 Hour Tolerance ppm of Calibrator Resistance pp...

Страница 61: ...olerance ppm of Calibrator Resistance ppm of Range 0 Ω 10 kΩ 0 23 0 22 0 Ω 1 kΩ 0 203 0 202 0 Ω 100 Ω 0 2010 0 2010 100 MΩ 100 MΩ 6000 40 900 20 10 MΩ 10 MΩ 400 10 100 2 1 MΩ 1 MΩ 90 10 20 2 100 kΩ 100 kΩ 80 8 15 4 10 kΩ 10 kΩ 80 23 12 22 1 kΩ 1 kΩ 80 203 12 202 100 Ω 100 Ω 80 2010 15 2010 Table 21 NI 4070 4072 DC Current Verification Limits Calibrator Amplitude Range 2 Year Limits Lower Upper 0 A...

Страница 62: ... Year Limits Amplitude Frequency Lower Upper 1 mA 1 kHz 10 mA 0 9976 mA 1 0024 mA 10 mA 1 kHz 10 mA 9 994 mA 10 006 mA 10 mA 1 kHz 100 mA 9 976 mA 10 024 mA 100 mA 1 kHz 100 mA 99 94 mA 100 06 mA 100 mA 1 kHz 1 A 99 7 mA 100 3 mA 1 A 1 kHz 1 A 0 9988 A 1 0012 A Table 23 Frequency Limits NI 6608 Output Frequency 2 Year Limits Lower Upper 1 Hz 0 9999 Hz 1 0001 Hz 20 kHz 19 998 kHz 20 002 kHz 500 kHz...

Страница 63: ...that correspond to the NI 4072 accuracy specifications Table 24 NI 4072 Capacitance and Inductance Verification Tolerances Verification Values Range 2 Year Tolerances of Reading of Range 0 pF 300 pF 0 0 5 0 μH 10 μH 0 1 270 pF 300 pF 0 15 0 5 1 nF 1 nF 0 15 0 1 1 nF 10 nF 0 15 0 1 100 nF 100 nF 0 15 0 1 100 nF 1 μF 0 18 0 1 10 μF 10 μF 0 18 0 1 10 μF 100 μF 0 18 0 1 1 000 μF 1 000 μF 0 18 0 1 1 00...

Страница 64: ...ccuracy limits or the 90 day accuracy limits if it has been externally calibrated within that time 2 Adjust the NI 4070 4072 3 Reverify the NI 4070 4072 using the 24 hour accuracy limits or the 2 year accuracy limits when the 24 hour limits are not specified Depending on your measurement and accuracy requirements a complete calibration of the NI 4070 4072 may not be necessary A number of options a...

Страница 65: ...Corporation 65 NI 4070 4072 Calibration Procedure Table 25 Summary of Adjustment Options Adjustment Optional Required DC Voltage Yes Resistance Yes AC Voltage Yes AC DC Current Yes Inductance and Capacitance NI 4072 only Yes ...

Страница 66: ...uency Mode Go to Verifying Capacitance and Inductance Go to Verifying Frequency Go to Verifying AC Current Go to Verifying DC Current Go to Verifying 2 Wire Resistance Go to Verifying 4 Wire Resistance Go to Verifying AC Voltage Go to Verifying DC Voltage Yes Yes Yes Yes Yes Yes Yes Yes No No No No No No No No Go to Initial Setup Pre Adjustment Post Adjustment Go to Adjustment Procedures Flowchart...

Страница 67: ...cation Procedures Flowchart Verification Go to Adjusting DC Voltage and Resistance Go to Adjusting AC Voltage Modes No No Yes Yes No Go to Completing the Adjustment Procedures Adjust AC Voltage Modes Adjust Current Modes Go to Adjusting Current Modes Perform Post Adjustment Verification Go to Adjusting Capacitance and Inductance ...

Страница 68: ...ruments Web site is your complete resource for technical support At ni com support you have access to everything from troubleshooting and application development self help resources to email and phone assistance from NI Application Engineers A Declaration of Conformity DoC is our claim of compliance with the Council of the European Communities using the manufacturer s declaration of conformity Thi...

Отзывы: