National Instruments Corporation
C-1
C
Common Questions
This appendix contains a list of common questions and answers relating to
the operation of the DAQCard-700. The questions are grouped according to
the type of information requested.
Installing and Configuring the DAQCard-700
Which NI documentation should I read first to get started using
NI-DAQ?
Refer to the
section of Chapter 2,
and Configuring the DAQCard-700
, for this information.
What version of NI-DAQ must I have to program the DAQCard-700?
The DAQCard-700 is supported by NI-DAQ for Windows versions 6.9.
x
and earlier. You may also use the DAQCard-700 with NI-DAQ for Mac OS
version 6.6.
x
or version 4.9.4 or earlier.
What is the best way to test the DAQCard without having to program
the DAQCard?
If you are using Windows, MAX has a Test Panel option that is available by
clicking
Devices and Interfaces
and then clicking the device. The Test
Panels are excellent tools for performing simple functional tests of the
DAQCard, such as analog input, digital I/O, and counter/timer tests. If you
are using Mac OS, the NI-DAQ Configuration Utility provides the same
functionality.
Operating the DAQCard-700
When using the DAQCard in differential analog input mode, I have
connected a differential input signal, but my readings are random and
drift rapidly. What’s wrong?
Check the ground-reference connections. The signal may be referenced to
a level that is considered
floating
with reference to the DAQCard ground
reference. Even if you are in DIFF input mode, the signal
must
still be