Chapter 4 Signal Connections
AT-MIO/AI E Series User Manual
4-12
National Instruments Corporation
Table 4-3 shows the I/O signal summary for the AT-MIO-16XE-10 and
AT-AI-16XE-10.
Table 4-3.
I/O Signal Summary, AT-MIO-16XE-10 and AT-AI-16XE-10
Signal Name
Drive
Impedance
Input/
Output
Protection
(Volts)
On/Off
Source
(mA at V)
Sink
(mA at
V)
Rise
Time
(ns)
Bias
ACH<0..15>
AI
100 G
Ω
in
parallel
with
100 pF
25/15
—
—
—
±
1 nA
AISENSE
AI
100 G
Ω
in
parallel
with
100 pF
25/15
—
—
—
±
1 nA
AIGND
AO
—
—
—
—
—
—
DAC0OUT
AO
0.1
Ω
Short-circuit
to ground
5 at 10
5 at -10
5
V/
µ
s
—
DAC1OUT
AO
0.1
Ω
Short-circuit
to ground
5 at 10
5 at -10
5
V/
µ
s
—
AOGND
AO
—
—
—
—
—
—
DGND
DO
—
—
—
—
—
—
VCC
DO
0.1
Ω
Short-circuit
to ground
1A
—
—
—
DIO<0..7>
DIO
—
Vcc +0.5
13 at (Vcc -0.4)
24 at 0.4
1.1
50 k
Ω
pu
SCANCLK
DO
—
—
3.5 at (Vcc -0.4)
5 at 0.4
1.5
50 k
Ω
pu
EXTSTROBE*
DO
—
—
3.5 at (Vcc -0.4)
5 at 0.4
1.5
50 k
Ω
pu
PFI0/TRIG1
DIO
—
Vcc +0.5
3.5 at (Vcc -0.4)
5 at 0.4
1.5
4.75 k
Ω
pu
PFI1/TRIG2
DIO
—
Vcc +0.5
3.5 at (Vcc -0.4)
5 at 0.4
1.5
50 k
Ω
pu
PFI2/CONVERT*
DIO
—
Vcc +0.5
3.5 at (Vcc -0.4)
5 at 0.4
1.5
50 k
Ω
pu
PFI3/GPCTR1_SOURCE
DIO
—
Vcc +0.5
3.5 at (Vcc -0.4)
5 at 0.4
1.5
50 k
Ω
pu
PFI4/GPCTR1_GATE
DIO
—
Vcc +0.5
3.5 at (Vcc -0.4)
5 at 0.4
1.5
50 k
Ω
pu
GPCTR1_OUT
DO
—
—
3.5 at (Vcc -0.4)
5 at 0.4
1.5
50 k
Ω
pu