©
National Instruments Corporation
45
NI 9219 Calibration Procedure
Full-Bridge Accuracy Adjustment
Complete the following steps to adjust the full-bridge accuracy of the NI 9219.
1.
Connect the NI 9219 to DMM1 as shown in Figure 15.
2.
Open a calibration session to your device using the DAQmx Initialize External Calibration VI. The
default password is
NI
.
3.
Set up the signal acquisition using the NI 9219 instance of the DAQmx Setup C Series
Calibration VI.
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxInitExtCal
with the following
parameters:
deviceName
: The name of the device you want to
calibrate. This name can be found under Devices
and Interfaces in Measurements & Automation
Explorer.
password
:
NI
calHandle
:
&calHandle
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxSetup9219Cal
with the following
parameters:
rangeMin
:
–0.0078125, –0.0625
rangeMax
:
0.0078125, 0.0625
calHandle
:
calHandle
channelName
:
cDAQ1Mod8/a
i
x
*
measType
:
DAQmx_Val_Bridge
bridgeConfig
:
DAQmx_Val_FullBridge
*
x
refers to the channel number