NI 9219 Calibration Procedure
6.
Obtain an array of the recommended calibration currents for your device using the NI 9219
instance of the DAQmx Get C Series Calibration Adjustment Points VI.
7.
Set the source to a reference value determined by the array of adjustment points.
8.
Configure the DMM for a current measurement.
9.
Enable Auto Zero on the DMM.
10. Acquire a current reading with the DMM.
11. Acquire a current reading with the NI 9219 using the NI 9219 instance of the DAQmx Adjust
C Series Calibration VI.
12. Repeat steps 7 through 11 for every value in the array.
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxGet9219CalAdjustPoints
with
the following parameters:
calHandle
:
calHandle
adjustmentPoints
: An array of reference values
for the DAQmx Adjust Calibration VI.
bufferSize
: The size of the
adjustmentPoints
array in elements.
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxAdjust9219Cal
with the following
parameters:
calHandle
:
calHandle
channelName
:
cDAQ1Mod8/a
i
x
*
value
: The DMM reading.
*
x
refers to the channel number