NI 9219 Calibration Procedure
32
ni.com
13. Compute and save the adjustment to the EEPROM using the DAQmx Close External Calibration
VI. This VI also saves the date, time, and temperature of the adjustment to the EEPROM.
14. Repeat steps 3 through 13 for each channel on the NI 9219.
15. Disconnect the source and the DMM from the device.
4-Wire and 2-Wire Resistance Accuracy Adjustment
Complete the following steps to adjust the 4-wire resistance accuracy and 2-wire resistance accuracy of
the NI 9219.
1.
Connect the NI 9219 to the resistor and the DMM as shown in Figure 7. Connect the sense leads
of the DMM as close as possible to the same point that the sense leads of the NI 9219 (HI and LO)
connect to the resistor to ensure that the DMM reference measurements include the same lead wire
resistance as the NI 9219 4-wire measurements.
2.
Open a calibration session to your device using the DAQmx Initialize External Calibration VI. The
default password is
NI
.
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxCloseExtCal
with the following
parameters:
calHandle
:
calHandle
action
:
DAQmx_Val_Action_Commit
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxInitExtCal
with the following
parameters:
deviceName
: The name of the device you want to
calibrate. This name can be found under Devices
and Interfaces in Measurement & Automation
Explorer.
password
:
NI
calHandle
:
&calHandle