Defibrillation

Chest Skin

About the Simulator…

The 

CR

i

S

is

System Defibrillation Chest

Skin has been designed to absorb a
maximum of 360 joules of energy*.
Although capable of absorbing 360
joules, we do recommend that the small-
est energy level possible be used while
training with the skin.

The 

CR

i

S

is

System Defibrillation Chest

Skin will enable you to practice defibril-
lation using manual, semi-automatic,
and automatic external defibrillators
(AEDs). When using any one of these in
training, always follow the recommended
operating procedures for that particular
defibrillator.

The 

CR

i

S

is

Manikin comes with the

Defibrillation Chest Skin in place and
ready for use.

*NOTE: 360 joules is the maximum energy
level that NASCO recommends administering
to the Defibrillation Chest Skin. Energy levels
in excess of 360 joules may cause irreparable
damage to the chest skin, circuitry, and
patient simulator being used — thus voiding

NASCO’s warranty and endangering your
equipment. NASCO assumes no liability for
damage or injury that may be caused by the
use and/or misuse of this equipment. All nor-
mal safety precautions for defibrillation train-
ing should be followed and energy levels
should be minimized. NASCO did not design
nor intend this Defibrillation Chest Skin to be
used as anything other than a training appa-
ratus for defibrillation.

Connecting Your Arrhythmia or
Patient Simulator:

Figure 1

NASCO has designed the 

CRiSis™

System to be compatible with a variety of
patient simulators. This is possible via
the standard four-lead snap cable
(Figure 1). If your patient simulator has
only two output posts, the red and black
leads must be connected to the patient
simulator.

You may want to connect the manikin to
the simulator that came with your defib-
rillator. If your patient simulator doesn’t
have the standard snap connectors, it
will be necessary to purchase the corre-
sponding adapters from NASCO. To
order, please see “Supplies/Replacement
Parts” on page 16.

2

The 

CR

i

S

is

™ 

Auscultation Manikin is a

Complete Resuscitation System that
allows training of CPR, airway manage-
ment, IV therapy, blood pressure read-
ings, defibrillation, and auscultation of
the heart and lung sounds.

This manual will guide you in setting up,
using, and maintaining each of the
available components. Each section also
includes a list of replacement parts, sup-
plies, and auxiliary equipment.

By reading and following all instructions 
carefully and completely, you can be
sure your 

Life/form

®

CR

i

S

is

Auscultation Manikin will provide years
of valuable service.

Cleaning:

Normal surface soil can be removed
from the trainer with mild soapy water.
Do not allow water to contact electrical
components. Stubborn stains may be
removed with REN Cleaner (W09919U).
Simply apply the REN Cleaner to the
soiled area and wipe clean with a soft
cloth. 

NOTE: Avoid using cleaner around the mouth
area if students will be applying direct mouth-
to-mouth resuscitation techniques, as the
cleaner may be toxic if ingested. NEVER place
the trainer on any kind of printed paper or
plastic. These materials, as well as ballpoint
pens, will transfer indelible stains. Do not use
any cosmetics. 

List of Components:

CR

i

S

is

Manikin

• Auscultation Chest
• SmartScope™
• Auscultation Chest Remote Control
• IV Arm
• Blood Pressure Arm with Speaker Wire
• Electronic Blood Pressure Control Unit
• Defibrillation Chest Skin
• 8 oz. Pump Spray Lubricant
• REN Cleaner
• 3cc Syringe with Needle
• 12cc Syringe with Needle
• 2 IV Bags
• 3 Pinch Clamps
• Small Towel (2)
• Butterfly Infusion Set
• Synthetic Blood
• 4 “AA” Batteries
• Sphygmomanometer

*Resusci

®

Anne™ is a trademark of Laerdal Medical Corporation.

15

Содержание Life/Form ADULT CRiSis LF03965U

Страница 1: ...h IV Trainer LF03000U CPARLENE Series LF03601U Adult Airway Management Trainer LF03602U Adult Airway Management on Manikin LF03603U Adult Airway Management Head Only LF03609U Child Airway Management Trainer LF03610U Child Airway Management Trainer Head Only LF03611U Child Defibrillation Chest Skin LF03612U Child IV Arm LF03613U Child Blood Pressure Arm LF03614U Child Intraosseous Infusion Femoral ...

Страница 2: ...nauthorized repair may void the warranty or cause further harm or damage to you or your equipment Supplies Replacement Parts for Defibrillation Chest Skin LF03656U Physio Control Defibrillation Pad and Patient Simulator Adapter Package LF03657U Marquette Electronics Defibrillation Pad and Patient Simulator Adapter Package LF03658U SpaceLabs First Medic or Laerdal Defibrillation Pad and Patient Sim...

Страница 3: ...nect the manikin to the simulator that came with your defib rillator If your patient simulator doesn t have the standard snap connectors it will be necessary to purchase the corre sponding adapters from NASCO To order please see Supplies Replacement Parts on page 16 2 The CRiSis Auscultation Manikin is a Complete Resuscitation System that allows training of CPR airway manage ment IV therapy blood ...

Страница 4: ... to damage the defibrillation electronics attached to the chest skin when removing the skin from the manikin For this example assume that the sounds stopped at 72 mmHg From the systolic window press the Menu key to change the diastolic window See figure 13 With the arrow down key set the correction to 2 Press the Menu key again and the message CALIBRATION COMPLETE will appear The main menu window ...

Страница 5: ...bsence of an audible pulse would have been noticed during what would have been the phase 2 Korotkoff sound It is this lack of an audible pulse that is consid ered an auscultatory gap Note that in reality the auscultatory gap can be pres ent in either the phase 1 or phase 2 Korotkoff sounds If it is desired to demonstrate the sounds heard while measuring a subject s blood pressure to the trainee or...

Страница 6: ...alloon and proceed Supplies Replacement Parts for Airway Management Trainer Head LF03285U Replacement Lungs LF03644U NASCO Pump Spray Lubricant W09919U REN Cleaner Pressing the Gap key simply turns the gap function off or on When the key is pressed a message will briefly appear that the ausculatory gap is enabled or disabled Also the main display will show at the bottom right of the display either...

Страница 7: ...eneral Instructions for Use A Preparing the Synthetic Blood 1 Fill the pint bottle containing synthet ic blood concentrate with distilled water Figure 3 2 Pour the synthetic blood into one of the bags Figure 4 Figure 3 Pinch Clamp Outlet Tubing Inlet Tubing Pinch Clamp 12 Life form Blood Venous System Anterior Posterior Needle Adapter Posterior Anterior Familiarizing Yourself with the NASCO Life f...

Страница 8: ... different Korotkoff phases can be identified and an optional auscultatory gap can be selected A pal pable radial pulse is present in the wrist For additional uses purchase the Blood Pressure Speaker System SB20146U for group demonstrations or review List of Components 1 Case 2 Arm with speaker wire 3 Electronic control unit 4 6 AA batteries 5 Sphygmomanometer General Instructions for Use Installi...

Страница 9: ...of infusion fluid from the IV bag B Control flow rate with clamp on IV set B This fluid can be used over If more real istic experience is desired with blood flashback instead of water when inserting butterfly into lumen of vein use next procedure C Figure 7 C Recommended Procedure for Simultaneous IV Infusions and Drawing Blood Using two IV bag kits hook up and install Figure 7 with IV bag A and I...

Страница 10: ...of infusion fluid from the IV bag B Control flow rate with clamp on IV set B This fluid can be used over If more real istic experience is desired with blood flashback instead of water when inserting butterfly into lumen of vein use next procedure C Figure 7 C Recommended Procedure for Simultaneous IV Infusions and Drawing Blood Using two IV bag kits hook up and install Figure 7 with IV bag A and I...

Страница 11: ... different Korotkoff phases can be identified and an optional auscultatory gap can be selected A pal pable radial pulse is present in the wrist For additional uses purchase the Blood Pressure Speaker System SB20146U for group demonstrations or review List of Components 1 Case 2 Arm with speaker wire 3 Electronic control unit 4 6 AA batteries 5 Sphygmomanometer General Instructions for Use Installi...

Страница 12: ...eneral Instructions for Use A Preparing the Synthetic Blood 1 Fill the pint bottle containing synthet ic blood concentrate with distilled water Figure 3 2 Pour the synthetic blood into one of the bags Figure 4 Figure 3 Pinch Clamp Outlet Tubing Inlet Tubing Pinch Clamp 12 Life form Blood Venous System Anterior Posterior Needle Adapter Posterior Anterior Familiarizing Yourself with the NASCO Life f...

Страница 13: ...alloon and proceed Supplies Replacement Parts for Airway Management Trainer Head LF03285U Replacement Lungs LF03644U NASCO Pump Spray Lubricant W09919U REN Cleaner Pressing the Gap key simply turns the gap function off or on When the key is pressed a message will briefly appear that the ausculatory gap is enabled or disabled Also the main display will show at the bottom right of the display either...

Страница 14: ...bsence of an audible pulse would have been noticed during what would have been the phase 2 Korotkoff sound It is this lack of an audible pulse that is consid ered an auscultatory gap Note that in reality the auscultatory gap can be pres ent in either the phase 1 or phase 2 Korotkoff sounds If it is desired to demonstrate the sounds heard while measuring a subject s blood pressure to the trainee or...

Страница 15: ... to damage the defibrillation electronics attached to the chest skin when removing the skin from the manikin For this example assume that the sounds stopped at 72 mmHg From the systolic window press the Menu key to change the diastolic window See figure 13 With the arrow down key set the correction to 2 Press the Menu key again and the message CALIBRATION COMPLETE will appear The main menu window ...

Страница 16: ...nect the manikin to the simulator that came with your defib rillator If your patient simulator doesn t have the standard snap connectors it will be necessary to purchase the corre sponding adapters from NASCO To order please see Supplies Replacement Parts on page 16 2 The CRiSis Auscultation Manikin is a Complete Resuscitation System that allows training of CPR airway manage ment IV therapy blood ...

Страница 17: ...nauthorized repair may void the warranty or cause further harm or damage to you or your equipment Supplies Replacement Parts for Defibrillation Chest Skin LF03656U Physio Control Defibrillation Pad and Patient Simulator Adapter Package LF03657U Marquette Electronics Defibrillation Pad and Patient Simulator Adapter Package LF03658U SpaceLabs First Medic or Laerdal Defibrillation Pad and Patient Sim...

Страница 18: ... the scroll button After the condition is selected press the enter button to activate The remote control can be programmed to these heart conditions 01 Normal 02 Aortic regurgitation 03 Pulmonary stenosis 04 Mitral stenosis 05 Holosystolic 06 Mid systolic 07 S3 Gallop 08 S4 Gallop 09 Systolic click 10 Atrial septal defect 11 PDA 12 VSD The remote control can be programmed to these lung conditions ...

Страница 19: ...h IV Trainer LF03000U CPARLENE Series LF03601U Adult Airway Management Trainer LF03602U Adult Airway Management on Manikin LF03603U Adult Airway Management Head Only LF03609U Child Airway Management Trainer LF03610U Child Airway Management Trainer Head Only LF03611U Child Defibrillation Chest Skin LF03612U Child IV Arm LF03613U Child Blood Pressure Arm LF03614U Child Intraosseous Infusion Femoral ...

Отзывы: