Nasco Life/Form ADULT CRiSis LF03965U Скачать руководство пользователя страница 11

10

7

Figure 4

3.  Be sure the clamp on the IV tubing is

closed, and hang the bag no more
than 18" above the level of the arm.

4.  Attach the end of the IV tubing to

one of the shoulder tubings.

5.  With the other shoulder tubing in a

basin or sink, gradually “flush” the
vascular system with synthetic blood
by slowly opening the clamp. Allow
some “blood” to pass through the
system until the air bubbles have
been eliminated.

6.  Once the system is filled, use one of

the pinch clamps to close off the
blood outlet tubing. The venous sys-
tem is now full of “blood” and pres-
surized. Be sure to leave the clamp
on the IV tubing open.

7.  After filling the venous system

according to instructions, the arm is
now ready for you to practice draw-
ing blood. Blood can be drawn any-
where along the pathway of the
vein. Distilled water, rather than
alcohol, should be used to prepare
the sites. Synthetic blood will actually
be aspirated once the vein is proper-
ly punctured.

8. Small diameter needles (20- to 25-

gauge) should be used.

Figure 5

B. Preparing the Arm for 
Intravenous Infusions

1.  Close the clamp at the end of IV

bag A tube, then fill with water (dis-
tilled water is recommended), and
hang not more than 18" above the
arm (Figure 5).

2.  Appropriate intravenous infusion

needles (or butterflies) should be
used, and distilled water is recom-
mended as an infusion.

3.  IVs can be started anywhere along

the pathway of the simulated vein.
Cleanse the sites with distilled water
only.

4.  Attach the adapter end of the IV tub-

ing into one of the shoulder tubing
ends.

5.  Place the other shoulder tubing end

in a basin or jar, and “flush” the vas-
cular system by opening the clamp.
Allow infusion (water) to pass
through the system until air bubbles
are eliminated. Shut off the flow with
a pinch clamp. The venous system is
now full and pressurized.

6.  Insert an IV needle or butterfly in the

vein. “Flashback” will indicate prop-
er insertion.

7.  Close the clamp on IV bag A tube

and remove pinch clamp from
shoulder tubing.

A

B

NASCO 

Life/form

®

Blood Pressure Simulator
About the Simulator…

The NASCO 

Life/form

®

Blood Pressure

Simulator is designed for years of mainte-
nance free operation as a training tool
for not only the nurse, doctor, or pre-hos-
pital healthcare provider, but also for
anyone involved in the training of health
care professionals.

The NASCO 

Life/form

®

Blood Pressure

Simulator has digitally recorded blood
pressure sounds that can be varied by pulse
rate and volume. The different Korotkoff
phases can be identified and an optional
auscultatory gap can be selected. A pal-
pable radial pulse is present in the wrist.
For additional uses, purchase the Blood
Pressure Speaker System (SB20146U) for
group demonstrations or review.

List of Components

1. Case
2. Arm with speaker wire
3. Electronic control unit
4. 6 “AA” batteries
5. Sphygmomanometer

General Instructions for Use:
Installing the Batteries

Take the Blood Pressure Electronic

Control Unit from the box and turn it over,
placing it face down onto a padded work
surface. Locate the “Open” compartment
on the back of the panel where the bat-
teries are to be installed. 

(See figure

1.)

Place your thumb or index finger on

the “Open” compartment and push up.

This will open the battery compart-

ment. The compartment is marked as to
the “+” and “–” positions of the batteries.
The battery bracket is now accessible to
the user.

Install 6 “AA” batteries as indicated by

the orientation diagram embossed in the
bottom of the bracket. It is recommended
that alkaline batteries be used for
increased battery life. After the batteries
have been properly installed, reassemble
the Blood Pressure Simulator by simply
reversing the disassembly procedures.
Place the unit face up on the work surface
and turn it on by pressing the on/off switch
on the top right of the unit. 

(See figure

2.)

Observe the display and verify that a

readable display is present, indicating prop-
er battery installation.

Note:

The control box has a battery

saving feature which will turn the unit off
after about 8-10 minutes if no keys are
used within that period of time.

Figure 2

The next step is to connect the simulat-

ed arm and speaker assembly along with
the sphygmomanometer cuff and gauge
assembly included with the unit. First,
locate the end of the pressure line attached
to the sphygmomanometer that has the
female luer fitting attached to it. Attach this
to the male luer fitting at the top of the unit
marked CUFF. 

(See figure 3.)

After the

pressure line fitting has been properly
installed, locate the plug that is at the end
of the wire which extends from the simulat-
ed arm assembly. Insert the plug from the
arm into the jack at the top of the unit
marked ARM. 

(See figure 4.) 

At this point the Blood Pressure Simulator
is ready for use. The unit has been facto-
ry calibrated for use with the accessories
included in the kit. No further calibration
adjustments should be neccessary at this
time. If the unit is to be used with a
sphygmomanometer other than that sup-
plied with the unit, or if recalibration is
necessary at a later date, then see the
section titled “Calibration Procedures”.

ON/OFF
SWITCH

Figure 1

Содержание Life/Form ADULT CRiSis LF03965U

Страница 1: ...h IV Trainer LF03000U CPARLENE Series LF03601U Adult Airway Management Trainer LF03602U Adult Airway Management on Manikin LF03603U Adult Airway Management Head Only LF03609U Child Airway Management Trainer LF03610U Child Airway Management Trainer Head Only LF03611U Child Defibrillation Chest Skin LF03612U Child IV Arm LF03613U Child Blood Pressure Arm LF03614U Child Intraosseous Infusion Femoral ...

Страница 2: ...nauthorized repair may void the warranty or cause further harm or damage to you or your equipment Supplies Replacement Parts for Defibrillation Chest Skin LF03656U Physio Control Defibrillation Pad and Patient Simulator Adapter Package LF03657U Marquette Electronics Defibrillation Pad and Patient Simulator Adapter Package LF03658U SpaceLabs First Medic or Laerdal Defibrillation Pad and Patient Sim...

Страница 3: ...nect the manikin to the simulator that came with your defib rillator If your patient simulator doesn t have the standard snap connectors it will be necessary to purchase the corre sponding adapters from NASCO To order please see Supplies Replacement Parts on page 16 2 The CRiSis Auscultation Manikin is a Complete Resuscitation System that allows training of CPR airway manage ment IV therapy blood ...

Страница 4: ... to damage the defibrillation electronics attached to the chest skin when removing the skin from the manikin For this example assume that the sounds stopped at 72 mmHg From the systolic window press the Menu key to change the diastolic window See figure 13 With the arrow down key set the correction to 2 Press the Menu key again and the message CALIBRATION COMPLETE will appear The main menu window ...

Страница 5: ...bsence of an audible pulse would have been noticed during what would have been the phase 2 Korotkoff sound It is this lack of an audible pulse that is consid ered an auscultatory gap Note that in reality the auscultatory gap can be pres ent in either the phase 1 or phase 2 Korotkoff sounds If it is desired to demonstrate the sounds heard while measuring a subject s blood pressure to the trainee or...

Страница 6: ...alloon and proceed Supplies Replacement Parts for Airway Management Trainer Head LF03285U Replacement Lungs LF03644U NASCO Pump Spray Lubricant W09919U REN Cleaner Pressing the Gap key simply turns the gap function off or on When the key is pressed a message will briefly appear that the ausculatory gap is enabled or disabled Also the main display will show at the bottom right of the display either...

Страница 7: ...eneral Instructions for Use A Preparing the Synthetic Blood 1 Fill the pint bottle containing synthet ic blood concentrate with distilled water Figure 3 2 Pour the synthetic blood into one of the bags Figure 4 Figure 3 Pinch Clamp Outlet Tubing Inlet Tubing Pinch Clamp 12 Life form Blood Venous System Anterior Posterior Needle Adapter Posterior Anterior Familiarizing Yourself with the NASCO Life f...

Страница 8: ... different Korotkoff phases can be identified and an optional auscultatory gap can be selected A pal pable radial pulse is present in the wrist For additional uses purchase the Blood Pressure Speaker System SB20146U for group demonstrations or review List of Components 1 Case 2 Arm with speaker wire 3 Electronic control unit 4 6 AA batteries 5 Sphygmomanometer General Instructions for Use Installi...

Страница 9: ...of infusion fluid from the IV bag B Control flow rate with clamp on IV set B This fluid can be used over If more real istic experience is desired with blood flashback instead of water when inserting butterfly into lumen of vein use next procedure C Figure 7 C Recommended Procedure for Simultaneous IV Infusions and Drawing Blood Using two IV bag kits hook up and install Figure 7 with IV bag A and I...

Страница 10: ...of infusion fluid from the IV bag B Control flow rate with clamp on IV set B This fluid can be used over If more real istic experience is desired with blood flashback instead of water when inserting butterfly into lumen of vein use next procedure C Figure 7 C Recommended Procedure for Simultaneous IV Infusions and Drawing Blood Using two IV bag kits hook up and install Figure 7 with IV bag A and I...

Страница 11: ... different Korotkoff phases can be identified and an optional auscultatory gap can be selected A pal pable radial pulse is present in the wrist For additional uses purchase the Blood Pressure Speaker System SB20146U for group demonstrations or review List of Components 1 Case 2 Arm with speaker wire 3 Electronic control unit 4 6 AA batteries 5 Sphygmomanometer General Instructions for Use Installi...

Страница 12: ...eneral Instructions for Use A Preparing the Synthetic Blood 1 Fill the pint bottle containing synthet ic blood concentrate with distilled water Figure 3 2 Pour the synthetic blood into one of the bags Figure 4 Figure 3 Pinch Clamp Outlet Tubing Inlet Tubing Pinch Clamp 12 Life form Blood Venous System Anterior Posterior Needle Adapter Posterior Anterior Familiarizing Yourself with the NASCO Life f...

Страница 13: ...alloon and proceed Supplies Replacement Parts for Airway Management Trainer Head LF03285U Replacement Lungs LF03644U NASCO Pump Spray Lubricant W09919U REN Cleaner Pressing the Gap key simply turns the gap function off or on When the key is pressed a message will briefly appear that the ausculatory gap is enabled or disabled Also the main display will show at the bottom right of the display either...

Страница 14: ...bsence of an audible pulse would have been noticed during what would have been the phase 2 Korotkoff sound It is this lack of an audible pulse that is consid ered an auscultatory gap Note that in reality the auscultatory gap can be pres ent in either the phase 1 or phase 2 Korotkoff sounds If it is desired to demonstrate the sounds heard while measuring a subject s blood pressure to the trainee or...

Страница 15: ... to damage the defibrillation electronics attached to the chest skin when removing the skin from the manikin For this example assume that the sounds stopped at 72 mmHg From the systolic window press the Menu key to change the diastolic window See figure 13 With the arrow down key set the correction to 2 Press the Menu key again and the message CALIBRATION COMPLETE will appear The main menu window ...

Страница 16: ...nect the manikin to the simulator that came with your defib rillator If your patient simulator doesn t have the standard snap connectors it will be necessary to purchase the corre sponding adapters from NASCO To order please see Supplies Replacement Parts on page 16 2 The CRiSis Auscultation Manikin is a Complete Resuscitation System that allows training of CPR airway manage ment IV therapy blood ...

Страница 17: ...nauthorized repair may void the warranty or cause further harm or damage to you or your equipment Supplies Replacement Parts for Defibrillation Chest Skin LF03656U Physio Control Defibrillation Pad and Patient Simulator Adapter Package LF03657U Marquette Electronics Defibrillation Pad and Patient Simulator Adapter Package LF03658U SpaceLabs First Medic or Laerdal Defibrillation Pad and Patient Sim...

Страница 18: ... the scroll button After the condition is selected press the enter button to activate The remote control can be programmed to these heart conditions 01 Normal 02 Aortic regurgitation 03 Pulmonary stenosis 04 Mitral stenosis 05 Holosystolic 06 Mid systolic 07 S3 Gallop 08 S4 Gallop 09 Systolic click 10 Atrial septal defect 11 PDA 12 VSD The remote control can be programmed to these lung conditions ...

Страница 19: ...h IV Trainer LF03000U CPARLENE Series LF03601U Adult Airway Management Trainer LF03602U Adult Airway Management on Manikin LF03603U Adult Airway Management Head Only LF03609U Child Airway Management Trainer LF03610U Child Airway Management Trainer Head Only LF03611U Child Defibrillation Chest Skin LF03612U Child IV Arm LF03613U Child Blood Pressure Arm LF03614U Child Intraosseous Infusion Femoral ...

Отзывы: