Nasco Life/Form ADULT CRiSis LF03965U Скачать руководство пользователя страница 18

About the Auscultation Simulator:

The 

Life/form

®

Auscultation Trainer Simulator

duplicates heart and lung conditions selected
by the instructor by wireless remote control.
The student must palpate to identify the correct
auscultation sites and will hear different heart
and lung sounds as the SmartScope™ is
moved from site to site. 

The simulator has six heart sites and five lung
sound sites on the anterior surface and
midaxillary sites. The remote control can
select from 12 different heart conditions, as
well as 16 lung conditions. The instructor can
select any condition then switch to another
condition easily so the student can compare
sounds and make a diagnosis.

The remote control does not have to be point-
ed directly at the manikin or stethoscope to
operate. One remote control will operate mul-
tiple sets of SmartScopes™ and manikins simul-
taneously. Great for group instruction. The
range of the remote control is up to 100 feet.

General Instructions for Use

To  begin using the Auscultation Trainer, press
the red power button. This turns on the remote
control and sends a signal to activate the
stethoscope as well. After the unit is activated,
the display will be in the “status” mode, display-
ing the current menu settings for the heart and
lung conditions.

To  select a new condition, press either the
heart or lung button. This will put the display
into the menu mode. The user can select a
condition by either using the number buttons,
or by viewing the conditions in sequence
using the scroll button. After the condition is
selected press the enter button to activate.

The remote control can be programmed to
these heart conditions:
01 Normal
02 Aortic regurgitation
03 Pulmonary stenosis
04 Mitral stenosis
05 Holosystolic
06 Mid-systolic
07 S3 Gallop
08 S4 Gallop
09 Systolic click
10 Atrial septal defect
11 PDA
12 VSD

The remote control can be programmed to
these lung conditions:
01 Normal Lung
02 Normal Vesicular
03 Wheezes
04 Mono wheeze
05 Fine crackle
06 Coarse crackle
07 Ronchi crackle
08 Stridor
09 Cavernous
10 Bronchovesicular
11 Bronchial 
12 Pulmonary edema
13 Infant
14 Friction rub
15 Egophony
16 Pectoriloquy

Consult the enclosed laminated card for
lung and heart sound site locations. 

To  listen to the selected sounds, place the
earpieces in ears angled in a forward posi-
tion. Place the diaphragm of the
SmartScope™over one of the appropriate
sites on the manikin (see diagram on back). 

As an alternative, an amplified speaker
(SB20146U) can be used allowing an entire
classroom to hear at the same time. To con-
nect the speaker, plug the speaker cord into
the speaker jack on the top of the Smart
Scope™ box. When a speaker is connected
to the SmartScope™, the ear pieces will not
work. Place the diaphragm of the
SmartScope™over the appropriate sites on
the manikin.

Note:

The remote has a battery saver that

shuts the unit down after 8 minutes if the
remote is left on the same setting. To
prevent it from shutting down, select a differ-
ent heart or lung sound within the 8 minute
period of time. 

Available Supplies 

LF01144U  Additional SmartScope™ 
LF01145U  Additional Auscultation 

Manikin Torso

LF01148U  Replacement Remote Control
SB20146U  Amplified Speaker

17

Содержание Life/Form ADULT CRiSis LF03965U

Страница 1: ...h IV Trainer LF03000U CPARLENE Series LF03601U Adult Airway Management Trainer LF03602U Adult Airway Management on Manikin LF03603U Adult Airway Management Head Only LF03609U Child Airway Management Trainer LF03610U Child Airway Management Trainer Head Only LF03611U Child Defibrillation Chest Skin LF03612U Child IV Arm LF03613U Child Blood Pressure Arm LF03614U Child Intraosseous Infusion Femoral ...

Страница 2: ...nauthorized repair may void the warranty or cause further harm or damage to you or your equipment Supplies Replacement Parts for Defibrillation Chest Skin LF03656U Physio Control Defibrillation Pad and Patient Simulator Adapter Package LF03657U Marquette Electronics Defibrillation Pad and Patient Simulator Adapter Package LF03658U SpaceLabs First Medic or Laerdal Defibrillation Pad and Patient Sim...

Страница 3: ...nect the manikin to the simulator that came with your defib rillator If your patient simulator doesn t have the standard snap connectors it will be necessary to purchase the corre sponding adapters from NASCO To order please see Supplies Replacement Parts on page 16 2 The CRiSis Auscultation Manikin is a Complete Resuscitation System that allows training of CPR airway manage ment IV therapy blood ...

Страница 4: ... to damage the defibrillation electronics attached to the chest skin when removing the skin from the manikin For this example assume that the sounds stopped at 72 mmHg From the systolic window press the Menu key to change the diastolic window See figure 13 With the arrow down key set the correction to 2 Press the Menu key again and the message CALIBRATION COMPLETE will appear The main menu window ...

Страница 5: ...bsence of an audible pulse would have been noticed during what would have been the phase 2 Korotkoff sound It is this lack of an audible pulse that is consid ered an auscultatory gap Note that in reality the auscultatory gap can be pres ent in either the phase 1 or phase 2 Korotkoff sounds If it is desired to demonstrate the sounds heard while measuring a subject s blood pressure to the trainee or...

Страница 6: ...alloon and proceed Supplies Replacement Parts for Airway Management Trainer Head LF03285U Replacement Lungs LF03644U NASCO Pump Spray Lubricant W09919U REN Cleaner Pressing the Gap key simply turns the gap function off or on When the key is pressed a message will briefly appear that the ausculatory gap is enabled or disabled Also the main display will show at the bottom right of the display either...

Страница 7: ...eneral Instructions for Use A Preparing the Synthetic Blood 1 Fill the pint bottle containing synthet ic blood concentrate with distilled water Figure 3 2 Pour the synthetic blood into one of the bags Figure 4 Figure 3 Pinch Clamp Outlet Tubing Inlet Tubing Pinch Clamp 12 Life form Blood Venous System Anterior Posterior Needle Adapter Posterior Anterior Familiarizing Yourself with the NASCO Life f...

Страница 8: ... different Korotkoff phases can be identified and an optional auscultatory gap can be selected A pal pable radial pulse is present in the wrist For additional uses purchase the Blood Pressure Speaker System SB20146U for group demonstrations or review List of Components 1 Case 2 Arm with speaker wire 3 Electronic control unit 4 6 AA batteries 5 Sphygmomanometer General Instructions for Use Installi...

Страница 9: ...of infusion fluid from the IV bag B Control flow rate with clamp on IV set B This fluid can be used over If more real istic experience is desired with blood flashback instead of water when inserting butterfly into lumen of vein use next procedure C Figure 7 C Recommended Procedure for Simultaneous IV Infusions and Drawing Blood Using two IV bag kits hook up and install Figure 7 with IV bag A and I...

Страница 10: ...of infusion fluid from the IV bag B Control flow rate with clamp on IV set B This fluid can be used over If more real istic experience is desired with blood flashback instead of water when inserting butterfly into lumen of vein use next procedure C Figure 7 C Recommended Procedure for Simultaneous IV Infusions and Drawing Blood Using two IV bag kits hook up and install Figure 7 with IV bag A and I...

Страница 11: ... different Korotkoff phases can be identified and an optional auscultatory gap can be selected A pal pable radial pulse is present in the wrist For additional uses purchase the Blood Pressure Speaker System SB20146U for group demonstrations or review List of Components 1 Case 2 Arm with speaker wire 3 Electronic control unit 4 6 AA batteries 5 Sphygmomanometer General Instructions for Use Installi...

Страница 12: ...eneral Instructions for Use A Preparing the Synthetic Blood 1 Fill the pint bottle containing synthet ic blood concentrate with distilled water Figure 3 2 Pour the synthetic blood into one of the bags Figure 4 Figure 3 Pinch Clamp Outlet Tubing Inlet Tubing Pinch Clamp 12 Life form Blood Venous System Anterior Posterior Needle Adapter Posterior Anterior Familiarizing Yourself with the NASCO Life f...

Страница 13: ...alloon and proceed Supplies Replacement Parts for Airway Management Trainer Head LF03285U Replacement Lungs LF03644U NASCO Pump Spray Lubricant W09919U REN Cleaner Pressing the Gap key simply turns the gap function off or on When the key is pressed a message will briefly appear that the ausculatory gap is enabled or disabled Also the main display will show at the bottom right of the display either...

Страница 14: ...bsence of an audible pulse would have been noticed during what would have been the phase 2 Korotkoff sound It is this lack of an audible pulse that is consid ered an auscultatory gap Note that in reality the auscultatory gap can be pres ent in either the phase 1 or phase 2 Korotkoff sounds If it is desired to demonstrate the sounds heard while measuring a subject s blood pressure to the trainee or...

Страница 15: ... to damage the defibrillation electronics attached to the chest skin when removing the skin from the manikin For this example assume that the sounds stopped at 72 mmHg From the systolic window press the Menu key to change the diastolic window See figure 13 With the arrow down key set the correction to 2 Press the Menu key again and the message CALIBRATION COMPLETE will appear The main menu window ...

Страница 16: ...nect the manikin to the simulator that came with your defib rillator If your patient simulator doesn t have the standard snap connectors it will be necessary to purchase the corre sponding adapters from NASCO To order please see Supplies Replacement Parts on page 16 2 The CRiSis Auscultation Manikin is a Complete Resuscitation System that allows training of CPR airway manage ment IV therapy blood ...

Страница 17: ...nauthorized repair may void the warranty or cause further harm or damage to you or your equipment Supplies Replacement Parts for Defibrillation Chest Skin LF03656U Physio Control Defibrillation Pad and Patient Simulator Adapter Package LF03657U Marquette Electronics Defibrillation Pad and Patient Simulator Adapter Package LF03658U SpaceLabs First Medic or Laerdal Defibrillation Pad and Patient Sim...

Страница 18: ... the scroll button After the condition is selected press the enter button to activate The remote control can be programmed to these heart conditions 01 Normal 02 Aortic regurgitation 03 Pulmonary stenosis 04 Mitral stenosis 05 Holosystolic 06 Mid systolic 07 S3 Gallop 08 S4 Gallop 09 Systolic click 10 Atrial septal defect 11 PDA 12 VSD The remote control can be programmed to these lung conditions ...

Страница 19: ...h IV Trainer LF03000U CPARLENE Series LF03601U Adult Airway Management Trainer LF03602U Adult Airway Management on Manikin LF03603U Adult Airway Management Head Only LF03609U Child Airway Management Trainer LF03610U Child Airway Management Trainer Head Only LF03611U Child Defibrillation Chest Skin LF03612U Child IV Arm LF03613U Child Blood Pressure Arm LF03614U Child Intraosseous Infusion Femoral ...

Отзывы: