3

NASCO 

Life/form

®

“Airway Larry”

Airway Management Head

About the Simulator…

The 

Life/form

®

“Airway Larry” Airway

Management Trainer is the most realistic
simulator available for the training of
intubation and other airway manage-
ment skills.

NASCO has taken great care to create
an airway management trainer that is
anatomically correct in respect to both
size and detail. Landmarks include: nos-
trils, teeth, tongue, oral and nasal phar-
ynx, larynx, cricoid ring, epiglottis, ary-
tenoid, false and true cords, trachea,
esophagus, “Airway Larry” lung set, and
stomach.

NASCO’s “Airway Larry” allows you to
practice oral, digital, and nasal intuba-
tion. E.T., E.O.A., PTL

®

, and Combitube

®

insertion can all be practiced as well
(please see “Using the Combitube

®

”).

Suction techniques and proper cuff infla-
tion can also be performed and evaluat-
ed.

Lubricating the Airway Trainer
Head:

Lubricate both the simulator and supplies
being used with the NASCO pump spray
lubricant provided (Figures 1 & 2).

Figure 1

Figure 2

NOTE: NASCO recommends the use of
the provided lubricant or a similar veg-
etable-based lubricant for the Airway
Management Trainer head. The use of a
silicone-based lubricant may cause dam-
age to the simulator, thus voiding
NASCO’s warranty on the trainer.

Set Up:
A. Connecting the Airway

1. The manikin’s airway has been dis-

connected to prevent damage dur-
ing shipping. To connect the airway,
remove the chest skin, chest plate,
compression springs, and upper
compression plate.  

NOTE: Caution must be taken not to
damage the defibrillation electronics
attached to the chest skin when removing
the skin from the manikin.

For this example, assume that the sounds
stopped at 72 mmHg. From the systolic
window, press the Menu key to change
the diastolic window. 

(See figure 13.)

With the arrow down key, set the correction
to –2. Press the Menu key again and the
message “CALIBRATION COMPLETE”
will appear. The main menu window will
be displayed.

Preparing Your Equipment for
Use with the NASCO 

Life/form

®

Blood Pressure Simulator

To  adapt your sphygmomanometer

for use with the simulator, it is first
necessary to obtain the luer fitting and the
T-fitting included with the simulator, which
is also available through your NASCO
catalog sales office. Using a wire cutting
pliers or similar instrument, carefully cut
the pressure line of the sphygmomanome-
ter about 2" from the gauge. 

(See figure

14.)

Take the T-fitting and insert it between

the two ends of tubing that were previous-
ly cut. 

(See figure 15.)

Assemble the

portion of pressure line tubing over the
remaining barbed end on the T-fitting.

(See figure 16.)

Figure 12

Figure13

Figure14

Lastly, take the loose end of the tub-

ing and push the female luer fitting into
it. 

(See figure 17.)

Attach the female

luer fitting to the fitting marked CUFF at
the top of the unit. Once all of the neces-
sary connections have been properly
made, go to the section titled “Calibration
Procedures” and calibrate the unit. When
the calibration procedures have been
completed, the simulator is ready for use.

Figure 15

Supplies/Replacement Parts for
Blood Pressure Simulator
LF01096U

Electronic Control Unit with

Sphygmomanometer

SB20146U

Blood Pressure Speaker

System

Figure 16

Figure 17

Содержание Life/Form ADULT CRiSis LF03965U

Страница 1: ...h IV Trainer LF03000U CPARLENE Series LF03601U Adult Airway Management Trainer LF03602U Adult Airway Management on Manikin LF03603U Adult Airway Management Head Only LF03609U Child Airway Management Trainer LF03610U Child Airway Management Trainer Head Only LF03611U Child Defibrillation Chest Skin LF03612U Child IV Arm LF03613U Child Blood Pressure Arm LF03614U Child Intraosseous Infusion Femoral ...

Страница 2: ...nauthorized repair may void the warranty or cause further harm or damage to you or your equipment Supplies Replacement Parts for Defibrillation Chest Skin LF03656U Physio Control Defibrillation Pad and Patient Simulator Adapter Package LF03657U Marquette Electronics Defibrillation Pad and Patient Simulator Adapter Package LF03658U SpaceLabs First Medic or Laerdal Defibrillation Pad and Patient Sim...

Страница 3: ...nect the manikin to the simulator that came with your defib rillator If your patient simulator doesn t have the standard snap connectors it will be necessary to purchase the corre sponding adapters from NASCO To order please see Supplies Replacement Parts on page 16 2 The CRiSis Auscultation Manikin is a Complete Resuscitation System that allows training of CPR airway manage ment IV therapy blood ...

Страница 4: ... to damage the defibrillation electronics attached to the chest skin when removing the skin from the manikin For this example assume that the sounds stopped at 72 mmHg From the systolic window press the Menu key to change the diastolic window See figure 13 With the arrow down key set the correction to 2 Press the Menu key again and the message CALIBRATION COMPLETE will appear The main menu window ...

Страница 5: ...bsence of an audible pulse would have been noticed during what would have been the phase 2 Korotkoff sound It is this lack of an audible pulse that is consid ered an auscultatory gap Note that in reality the auscultatory gap can be pres ent in either the phase 1 or phase 2 Korotkoff sounds If it is desired to demonstrate the sounds heard while measuring a subject s blood pressure to the trainee or...

Страница 6: ...alloon and proceed Supplies Replacement Parts for Airway Management Trainer Head LF03285U Replacement Lungs LF03644U NASCO Pump Spray Lubricant W09919U REN Cleaner Pressing the Gap key simply turns the gap function off or on When the key is pressed a message will briefly appear that the ausculatory gap is enabled or disabled Also the main display will show at the bottom right of the display either...

Страница 7: ...eneral Instructions for Use A Preparing the Synthetic Blood 1 Fill the pint bottle containing synthet ic blood concentrate with distilled water Figure 3 2 Pour the synthetic blood into one of the bags Figure 4 Figure 3 Pinch Clamp Outlet Tubing Inlet Tubing Pinch Clamp 12 Life form Blood Venous System Anterior Posterior Needle Adapter Posterior Anterior Familiarizing Yourself with the NASCO Life f...

Страница 8: ... different Korotkoff phases can be identified and an optional auscultatory gap can be selected A pal pable radial pulse is present in the wrist For additional uses purchase the Blood Pressure Speaker System SB20146U for group demonstrations or review List of Components 1 Case 2 Arm with speaker wire 3 Electronic control unit 4 6 AA batteries 5 Sphygmomanometer General Instructions for Use Installi...

Страница 9: ...of infusion fluid from the IV bag B Control flow rate with clamp on IV set B This fluid can be used over If more real istic experience is desired with blood flashback instead of water when inserting butterfly into lumen of vein use next procedure C Figure 7 C Recommended Procedure for Simultaneous IV Infusions and Drawing Blood Using two IV bag kits hook up and install Figure 7 with IV bag A and I...

Страница 10: ...of infusion fluid from the IV bag B Control flow rate with clamp on IV set B This fluid can be used over If more real istic experience is desired with blood flashback instead of water when inserting butterfly into lumen of vein use next procedure C Figure 7 C Recommended Procedure for Simultaneous IV Infusions and Drawing Blood Using two IV bag kits hook up and install Figure 7 with IV bag A and I...

Страница 11: ... different Korotkoff phases can be identified and an optional auscultatory gap can be selected A pal pable radial pulse is present in the wrist For additional uses purchase the Blood Pressure Speaker System SB20146U for group demonstrations or review List of Components 1 Case 2 Arm with speaker wire 3 Electronic control unit 4 6 AA batteries 5 Sphygmomanometer General Instructions for Use Installi...

Страница 12: ...eneral Instructions for Use A Preparing the Synthetic Blood 1 Fill the pint bottle containing synthet ic blood concentrate with distilled water Figure 3 2 Pour the synthetic blood into one of the bags Figure 4 Figure 3 Pinch Clamp Outlet Tubing Inlet Tubing Pinch Clamp 12 Life form Blood Venous System Anterior Posterior Needle Adapter Posterior Anterior Familiarizing Yourself with the NASCO Life f...

Страница 13: ...alloon and proceed Supplies Replacement Parts for Airway Management Trainer Head LF03285U Replacement Lungs LF03644U NASCO Pump Spray Lubricant W09919U REN Cleaner Pressing the Gap key simply turns the gap function off or on When the key is pressed a message will briefly appear that the ausculatory gap is enabled or disabled Also the main display will show at the bottom right of the display either...

Страница 14: ...bsence of an audible pulse would have been noticed during what would have been the phase 2 Korotkoff sound It is this lack of an audible pulse that is consid ered an auscultatory gap Note that in reality the auscultatory gap can be pres ent in either the phase 1 or phase 2 Korotkoff sounds If it is desired to demonstrate the sounds heard while measuring a subject s blood pressure to the trainee or...

Страница 15: ... to damage the defibrillation electronics attached to the chest skin when removing the skin from the manikin For this example assume that the sounds stopped at 72 mmHg From the systolic window press the Menu key to change the diastolic window See figure 13 With the arrow down key set the correction to 2 Press the Menu key again and the message CALIBRATION COMPLETE will appear The main menu window ...

Страница 16: ...nect the manikin to the simulator that came with your defib rillator If your patient simulator doesn t have the standard snap connectors it will be necessary to purchase the corre sponding adapters from NASCO To order please see Supplies Replacement Parts on page 16 2 The CRiSis Auscultation Manikin is a Complete Resuscitation System that allows training of CPR airway manage ment IV therapy blood ...

Страница 17: ...nauthorized repair may void the warranty or cause further harm or damage to you or your equipment Supplies Replacement Parts for Defibrillation Chest Skin LF03656U Physio Control Defibrillation Pad and Patient Simulator Adapter Package LF03657U Marquette Electronics Defibrillation Pad and Patient Simulator Adapter Package LF03658U SpaceLabs First Medic or Laerdal Defibrillation Pad and Patient Sim...

Страница 18: ... the scroll button After the condition is selected press the enter button to activate The remote control can be programmed to these heart conditions 01 Normal 02 Aortic regurgitation 03 Pulmonary stenosis 04 Mitral stenosis 05 Holosystolic 06 Mid systolic 07 S3 Gallop 08 S4 Gallop 09 Systolic click 10 Atrial septal defect 11 PDA 12 VSD The remote control can be programmed to these lung conditions ...

Страница 19: ...h IV Trainer LF03000U CPARLENE Series LF03601U Adult Airway Management Trainer LF03602U Adult Airway Management on Manikin LF03603U Adult Airway Management Head Only LF03609U Child Airway Management Trainer LF03610U Child Airway Management Trainer Head Only LF03611U Child Defibrillation Chest Skin LF03612U Child IV Arm LF03613U Child Blood Pressure Arm LF03614U Child Intraosseous Infusion Femoral ...

Отзывы: