Nasco Life/Form ADULT CRiSis LF03965U Скачать руководство пользователя страница 13

12

Figure 7

Next, take the trainer to an area with a
sink and open counter space. Using the
red caps supplied, plug off the right
bronchus (the esophagus should already
be plugged). Stabilize the head on the
counter face-up (towels work well for this)
with plugged tubes hanging over the sink
(Figure 8). Carefully pour warm soapy
water (a mild dish soap works best) into
the mouth until the water level reaches
halfway up the tongue. Tilt the head back
and bring the neck up 3" off the counter-
top.

Figure 8

Continue filling until the water level covers
the tongue. At this point, take a small soft
brush and gently scrub the inside of the
mouth (a small toothbrush works well for
this). Cotton swabs can be used to scrub
inside the nostrils. When done, pull the
plug from the esophagus and drain the
water into the sink. Now pick the head up
to a vertical position and pull the plugs
from the trachea to completely empty the
system (Figure 9).

Rinsing the Airway:

To rinse the airway, follow the same pro-
cedure using clean, warm tap water.
Repeat this process until all the soap has
been flushed from the system.

Figure 9

Disinfecting the Airway:

To disinfect, repeat the standard cleaning
procedure, but this time use a bleach
solution, as specified by the Centers for
Disease Control, instead of soapy water.
Fill the system with the solution until it
reaches the corners of the mouth.
Remember to start filling with the head
flat and finish with the neck slightly ele-
vated to ensure that the solution com-
pletely fills all airway passages. Once
completely filled with the bleach solu-
tion, allow the head to sit for at least 10
minutes. Drain as described earlier and
repeat the rinsing process to flush out all
of the bleach solution. Set the head
aside and allow it to dry completely.

Using the Combitube

®

:

Thoroughly read and follow the instruc-
tions that come with the Combitube

®

. The

trainer will accept either a full-size or a
small adult tube. As with a live patient, it
may be necessary to back the tube out
slightly if ventilation cannot be established.

NOTE: Depending on tube placement, the
large cuff may not accept the recommended
amount of air. In this case, simply inflate the
cuff to its maximum volume (when the plunger
stops) and, while holding the plunger down,
detach the syringe from the blue pilot balloon,
and proceed.

Supplies/Replacement Parts for
Airway Management Trainer Head:

LF03285U Replacement Lungs
LF03644U NASCO Pump Spray Lubricant
W09919U REN Cleaner

Pressing the Gap key simply turns the
gap function off or on. When the key is
pressed, a message will briefly appear
that the ausculatory gap is enabled or
disabled. Also the main display will show
(at the bottom right of the display) either
AGap:Y (for on) or AGap:N (for off).

The arrow up and down keys also

control the volume of the sounds that are
present in the arm. From the main menu,
press the up arrow key to increase the
volume, press the down arrow key to
decrease the volume. The volume levels
can be adjusted from level 1 (the lowest
volume) to level 7 (the highest volume).

Using the NASCO 

Life/form

®

Blood Pressure Simulator

First, verify that the pressure line tubing

from the sphygmomanometer and the
audio line coming from the simulated arm
assembly are properly connected to the
blood pressure simulator unit, as previous-
ly described in the set up procedures.
Apply the sphygmomanometer cuff and
gauge to the simulated arm assembly in
the usual manner. Apply the stethoscope to
the simulated arm also in the usual man-
ner. Set the systolic and diastolic controls
to the desired levels. Select the auscultato-
ry gap if desired. Finally, adjust the pulse
rate control to the desired setting.

Figure 8

Figure 9

Figure 11

Figure 10

A pulseless condition can be simulated

by switching your unit to the “pulseless”
mode. Press the Menu key repeatedly
until “Set PALPATION” menu appears. By
pressing the down arrow at this point,
palpations can be disabled causing the
simulator to be pulseless. When in the
“pulseless” mode, all settings are auto-
matically reset to 0 and all blood pres-
sure sounds are disabled.

The pulseless setting will also turn off

the sounds in the arm. The pulse will
always be on unless the pulseless feature
is activated or if the systolic or heart rate
levels are set to zero. To do this, press the
Menu key four times. The down arrow
key will set the pulse to pulseless. Press
the arrow up key to turn the pulse back
on. 

(See figure 10.)

Located to the

right of the Menu key is the ausculatory
Gap key. 

(See figure 11.) 

This key is

included to simulate the ausculatory gap
that is sometimes present between phase
1 and phase 2 sounds in which no audi-
ble sound is noted during this portion of
ausculation. This control function is includ-
ed so that the trainee can become famil-
iar with this phenomenon. 

Pulse Location

5

Содержание Life/Form ADULT CRiSis LF03965U

Страница 1: ...h IV Trainer LF03000U CPARLENE Series LF03601U Adult Airway Management Trainer LF03602U Adult Airway Management on Manikin LF03603U Adult Airway Management Head Only LF03609U Child Airway Management Trainer LF03610U Child Airway Management Trainer Head Only LF03611U Child Defibrillation Chest Skin LF03612U Child IV Arm LF03613U Child Blood Pressure Arm LF03614U Child Intraosseous Infusion Femoral ...

Страница 2: ...nauthorized repair may void the warranty or cause further harm or damage to you or your equipment Supplies Replacement Parts for Defibrillation Chest Skin LF03656U Physio Control Defibrillation Pad and Patient Simulator Adapter Package LF03657U Marquette Electronics Defibrillation Pad and Patient Simulator Adapter Package LF03658U SpaceLabs First Medic or Laerdal Defibrillation Pad and Patient Sim...

Страница 3: ...nect the manikin to the simulator that came with your defib rillator If your patient simulator doesn t have the standard snap connectors it will be necessary to purchase the corre sponding adapters from NASCO To order please see Supplies Replacement Parts on page 16 2 The CRiSis Auscultation Manikin is a Complete Resuscitation System that allows training of CPR airway manage ment IV therapy blood ...

Страница 4: ... to damage the defibrillation electronics attached to the chest skin when removing the skin from the manikin For this example assume that the sounds stopped at 72 mmHg From the systolic window press the Menu key to change the diastolic window See figure 13 With the arrow down key set the correction to 2 Press the Menu key again and the message CALIBRATION COMPLETE will appear The main menu window ...

Страница 5: ...bsence of an audible pulse would have been noticed during what would have been the phase 2 Korotkoff sound It is this lack of an audible pulse that is consid ered an auscultatory gap Note that in reality the auscultatory gap can be pres ent in either the phase 1 or phase 2 Korotkoff sounds If it is desired to demonstrate the sounds heard while measuring a subject s blood pressure to the trainee or...

Страница 6: ...alloon and proceed Supplies Replacement Parts for Airway Management Trainer Head LF03285U Replacement Lungs LF03644U NASCO Pump Spray Lubricant W09919U REN Cleaner Pressing the Gap key simply turns the gap function off or on When the key is pressed a message will briefly appear that the ausculatory gap is enabled or disabled Also the main display will show at the bottom right of the display either...

Страница 7: ...eneral Instructions for Use A Preparing the Synthetic Blood 1 Fill the pint bottle containing synthet ic blood concentrate with distilled water Figure 3 2 Pour the synthetic blood into one of the bags Figure 4 Figure 3 Pinch Clamp Outlet Tubing Inlet Tubing Pinch Clamp 12 Life form Blood Venous System Anterior Posterior Needle Adapter Posterior Anterior Familiarizing Yourself with the NASCO Life f...

Страница 8: ... different Korotkoff phases can be identified and an optional auscultatory gap can be selected A pal pable radial pulse is present in the wrist For additional uses purchase the Blood Pressure Speaker System SB20146U for group demonstrations or review List of Components 1 Case 2 Arm with speaker wire 3 Electronic control unit 4 6 AA batteries 5 Sphygmomanometer General Instructions for Use Installi...

Страница 9: ...of infusion fluid from the IV bag B Control flow rate with clamp on IV set B This fluid can be used over If more real istic experience is desired with blood flashback instead of water when inserting butterfly into lumen of vein use next procedure C Figure 7 C Recommended Procedure for Simultaneous IV Infusions and Drawing Blood Using two IV bag kits hook up and install Figure 7 with IV bag A and I...

Страница 10: ...of infusion fluid from the IV bag B Control flow rate with clamp on IV set B This fluid can be used over If more real istic experience is desired with blood flashback instead of water when inserting butterfly into lumen of vein use next procedure C Figure 7 C Recommended Procedure for Simultaneous IV Infusions and Drawing Blood Using two IV bag kits hook up and install Figure 7 with IV bag A and I...

Страница 11: ... different Korotkoff phases can be identified and an optional auscultatory gap can be selected A pal pable radial pulse is present in the wrist For additional uses purchase the Blood Pressure Speaker System SB20146U for group demonstrations or review List of Components 1 Case 2 Arm with speaker wire 3 Electronic control unit 4 6 AA batteries 5 Sphygmomanometer General Instructions for Use Installi...

Страница 12: ...eneral Instructions for Use A Preparing the Synthetic Blood 1 Fill the pint bottle containing synthet ic blood concentrate with distilled water Figure 3 2 Pour the synthetic blood into one of the bags Figure 4 Figure 3 Pinch Clamp Outlet Tubing Inlet Tubing Pinch Clamp 12 Life form Blood Venous System Anterior Posterior Needle Adapter Posterior Anterior Familiarizing Yourself with the NASCO Life f...

Страница 13: ...alloon and proceed Supplies Replacement Parts for Airway Management Trainer Head LF03285U Replacement Lungs LF03644U NASCO Pump Spray Lubricant W09919U REN Cleaner Pressing the Gap key simply turns the gap function off or on When the key is pressed a message will briefly appear that the ausculatory gap is enabled or disabled Also the main display will show at the bottom right of the display either...

Страница 14: ...bsence of an audible pulse would have been noticed during what would have been the phase 2 Korotkoff sound It is this lack of an audible pulse that is consid ered an auscultatory gap Note that in reality the auscultatory gap can be pres ent in either the phase 1 or phase 2 Korotkoff sounds If it is desired to demonstrate the sounds heard while measuring a subject s blood pressure to the trainee or...

Страница 15: ... to damage the defibrillation electronics attached to the chest skin when removing the skin from the manikin For this example assume that the sounds stopped at 72 mmHg From the systolic window press the Menu key to change the diastolic window See figure 13 With the arrow down key set the correction to 2 Press the Menu key again and the message CALIBRATION COMPLETE will appear The main menu window ...

Страница 16: ...nect the manikin to the simulator that came with your defib rillator If your patient simulator doesn t have the standard snap connectors it will be necessary to purchase the corre sponding adapters from NASCO To order please see Supplies Replacement Parts on page 16 2 The CRiSis Auscultation Manikin is a Complete Resuscitation System that allows training of CPR airway manage ment IV therapy blood ...

Страница 17: ...nauthorized repair may void the warranty or cause further harm or damage to you or your equipment Supplies Replacement Parts for Defibrillation Chest Skin LF03656U Physio Control Defibrillation Pad and Patient Simulator Adapter Package LF03657U Marquette Electronics Defibrillation Pad and Patient Simulator Adapter Package LF03658U SpaceLabs First Medic or Laerdal Defibrillation Pad and Patient Sim...

Страница 18: ... the scroll button After the condition is selected press the enter button to activate The remote control can be programmed to these heart conditions 01 Normal 02 Aortic regurgitation 03 Pulmonary stenosis 04 Mitral stenosis 05 Holosystolic 06 Mid systolic 07 S3 Gallop 08 S4 Gallop 09 Systolic click 10 Atrial septal defect 11 PDA 12 VSD The remote control can be programmed to these lung conditions ...

Страница 19: ...h IV Trainer LF03000U CPARLENE Series LF03601U Adult Airway Management Trainer LF03602U Adult Airway Management on Manikin LF03603U Adult Airway Management Head Only LF03609U Child Airway Management Trainer LF03610U Child Airway Management Trainer Head Only LF03611U Child Defibrillation Chest Skin LF03612U Child IV Arm LF03613U Child Blood Pressure Arm LF03614U Child Intraosseous Infusion Femoral ...

Отзывы: