Set Up
Injectable Arm
Care and Maintenance
After each use of the Injectable Training Arm, follow
these procedures:
1 . Disconnect IV bags, remove infusion needles, and
flush the venous system using distilled water and
12 cc syringe .
2 . Simulated blood can be returned to its bottle and
reused .
3 . Rinse IV bag containing simulated blood with
distilled water, flushing through tubing into a
sink or basin .
4 . Remove pinch clamps from IV bags and injectable
training arm shoulder tubing .
5 . Wash the outside of the injectable training arm
with mild liquid soap . Stubborn stains may be
washed with Nasco cleaner . Dispense Nasco
cleaner on clean, soft, dry cloth and gently wipe
soiled area .
6 . Remove excess water from the venous system
by raising the hand, lowering the shoulder, and
draining it into a sink or basin .
7 . Allow the arm to dry completely before storing .
User Help Guide for the Injectable Training Arm
1 . Follow procedures and ensure clamps are open
on appropriate fluid supply/IV bags and/or
shoulder tubes .
2 . Look over your equipment prior to use . IV tubes
will kink at the clamp site with repeated use .
Routinely move clamps up or down the tube to
reduce the probability of kinks . When kink occurs,
slide the clamp to a new position and, with fingers,
massage tubing at pinched site to restore lumen .
Replacement fluid bags are available . Removing
clamps prior to storage is recommended .
3 . Check to ensure hanging bags are hung to the
appropriate height . Hanging the supply bags
slightly higher for bags not producing enough
pressure can create just enough gravitational
force on the fluid to facilitate flow .
4 . If a venous system clog is suspected, try using
a large 50 cc syringe to force distilled water
through the tubing .
5 . Check the venous system tubing for kinks . First,
lubricate the outside of the arm skin generously
with mild liquid soap . Peel the skin back to the
knuckles, being careful NOT to remove the skin
from the fingers . Examine all the tubing for
possible kinks . Replace the skin and infiltrate the
system again .