Rev 8, Dated 27-2-2018
14
Problem
4.
On/Off switch (mcb) of treadmill trips down
Possible cause
:
Due to sudden large spike or high voltage, ac/dc drive may be short-
circuited.
Solution
:
1.
Isolate the treadmill from PC by removing treadmill to STWIN
amplifier cable. Switch on treadmill.
2.
If still mcb trips, change AC/DC drive.
Material
Req.
AC/DC Drive, silicon oil (200ml).
Fig 6. FRONT PANEL CONNECTIO
Problem
5.
Treadmill elevates properly up to 1
st
stage only and no further
elevation.
Possible cause:
Treadmill not calibrated.
Solution:
Calibrate the treadmill as per user manual.
10k 10 turn pot setting might get disturbed or it is faulty.
Material
Req.
10k 10 turn pot along with 33 t gear wheel, silicon oil (200ml).
FROM AC
MAINS
MCB
CHOKE
TO TPC CABLE
TO
EMERGEN
CY STOP
SWITCH
10 K 10
TURN
POT
AC MOTOR
CONNECTION
Содержание ST WIN
Страница 1: ...Rev 8 Dated 27 2 2018 1 ST WIN Service Manual Nasan Medical Electronics Pvt Ltd All Rights Reserved...
Страница 22: ...Rev 8 Dated 27 2 2018 22 Fig 9 WIRING DIAGRAM OF AMPLIFIER...
Страница 67: ...Rev 8 Dated 27 2 2018 67 4 Take 20 ml of silicon oil and pour from bottom side evenly...