Rev 8, Dated 27-2-2018
40
D) Restore settings from the floppy. --Now this feature is not used
Choose this option from
‘Utility’
menu to install the setting saved on the
floppy. When pressed this button, setting stored in the floppy are restored
in the newly installed software.
--E) Delete printouts from printer queue …
To delete printouts from printer queue go to Utility and select
‘Delete
printouts from printer queue. .
’Option.
‘Printer and Fax’ dialog
box
will be displayed. Select the printer which is connected to ST win
unit and double click on it. It will display that documents stuck in the
queue and if those documents in the queue are not required then right
click on it and select ‘cancel’ option. It will delete the documents in
queue.
H. Precautions:
1.
Precaution to be taken while testing the AC-DC drive
1
. Do not touch any part on dc drive except its cover during ON condition.
Otherwise you will experience a hazardous shock and the DC drive will be damaged.
2
. As the AC/DC drive is floating, do not connect earth or ground to it. This
will damage the dc motor circuit of the card.
3
. While interchanging AC/DC drive, always refer to wiring diagram.
2.
Earthing:
The Earthing for the system should be proper. This is to ensure patient safety
and for no interference of line voltage fluctuations. For correct earthing procedure,
refer to work order booklet and get it done and approved from licensed electrician
only.
If the earthing is not proper, leakage currents may pass from the amplifier to
ground through the patient. If the current magnitude is large, it can be harmful for
the patient.
3. If earth wire is showing 0 Volt, check whether it is connected
or not.
4. General precautions for using the treadmill-
Read this section before operating the treadmill. Keep the area underneath
the treadmill clear.
Do not start the treadmill when someone is standing on the belt.
Do not leave a patient unattended on the treadmill.
Do not stop the treadmill if someone is walking on it except in case of
emergency.
Keep speed and grade at minimum when patients are getting on the
treadmill
.
5.
Check voltage before replacing/touching any part.