Alternate Display Screen
The
Alternate Display Screen
is very similar to the
Main
Measurement Display Screen
with one difference. On the third
line an instantaneous indication of the percent of standard is
shown instead of displaying a maximum value. This screen
should only be used with flat response probes for single
frequency measurements or where all the emitter frequencies
are close and the standard has the same limits.
To use the Alternate Display Screen...
1.
Press
F4 (FREQ)
and then enter the frequency that
you are measuring. You must fill all four positions. Use
F1
,
F2
, or
F3
as required to change the units.
2.
Press
F4 (SAVE)
. The probe correction factor will be
displayed briefly if you are using a probe in memory
and you will return to the main measurement screen.
3.
Press
F2 (MENU)
, then
8 (NEXT)
to access the
second menu screen. Select
STNDS (6)
. Press the
appropriate standard that you want to reference. (You
must use a function key even if the current setting is
correct.) A new screen will appear.
4.
Press
F1
to verify or
F4
to make another selection.
Press
Esc
twice to go back to the main measurement
screen.
5.
From the Main Measurement Screen select
%STD
by
pressing
F3
. The meter will now automatically display
the instantaneous measured value in the chosen units
on the second line and at the same time compare that
FIELD STRENGTH
• Indicates actual
measured value
• Unit of measure can be
changed by going to the
First Menu Screen
. 2 6
W
m
m
c
/
2
1 2 5 . 2 %
S T D
1 0 . 0 0
H z
G
L O G
M E N U
M A X
F R E Q
6
F1
F2
F3
F4
CALCULATED
PERCENT OF
STANDARD
• Displays percent of
standard that has been
calculated from the
field strength reading
using the referenced
standard and the
indicated frequency